Introduction

It is important to understand the basic techniques of radiotherapy planning because these will help when developing complex plans. These techniques, which are taught at the First FRCR level (Royal College of Radiologists), are used by the treatment centre's Physics Department for checking the validity of calculations before they are applied to patients, and anyone interpreting plans will need to be familiar with them to know whether it is worth adjusting treatment plans during review. Advanced radiotherapy techniques will be discussed in Chapter 6.

There are several useful reviews of radiotherapy physics in the literature. One by Shiu and Mellenberg (2001) includes sections on isodose planning. Another, by Purdy (2000), provides a perspective on future directions in 3D treatment planning. Radiotherapy is a rapidly developing field and it is important to ensure that new methods satisfy safety and effectiveness requirements before being adopted as routine treatments.

This chapter will deal with the general principles of developing isodose plans that are suitable for treatment and will use specific examples to highlight particular points. It will focus mainly on external beam radiotherapy with megavoltage photons; there will be shorter sections on the use of electrons and brachytherapy. Some aspects of radiotherapy planning using lower energy (kilovoltage) photons will be discussed in Chapter 36. The use of unsealed radiation sources for prostate, thyroid and childhood cancers are discussed in Chapters 22, 38 and 40, respectively.

Treatment planning overview

Radiotherapy planning can be divided into stages as follows.

  1. • Patient preparation, position and immobilisation.

  2. • Localisation method (e.g. orthogonal films, CT scanning, and image co-registration).

  3. • Definition of target volumes and organs at risk.

  4. • Radiotherapy technique, including beam arrangements, beam energy, size and shape, weighting, wedges and production of isodose plan.

  5. • Prescription, including number of phases, dose, energy and fractionation.

  6. • Verification (i.e. checking the geometrical set-up of the treatment). This can take place before treatment in the simulator or CT simulator, and/or during treatment using portal imaging.

Quality assurance (QA) is essential and it is of paramount importance for each member of the planning team to be familiar with the overall process.

Patient position and immobilisation

For radiotherapy to be effective it must be delivered accurately, and it is important to understand the methods of immobilisation and the inevitable uncertainties in the delivery of treatment.

Type
Chapter
Information
Publisher: Cambridge University Press
Print publication year: 2015

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Apisarnthanarax, S., Elliott, D. D., El-Nagger, A. K., et al. (2006). Determining optimal clinical target volume margins in head-and-neck cancer based on microscopic extracapsular extension of metastatic neck nodes. Int. J. Radiat. Oncol. Biol. Phys., 64, 678–683.CrossRefGoogle ScholarPubMed
BIR. (2003). Geometric Uncertainties in Radiotherapy. Defining the Planning Target Volume. Prepared by a Working Party of the British Institute of Radiology. London: British Institute of Radiology.
Dearnaley, D. P., Khoo, V. S., Norman, A. R., et al. (1999). Comparison of radiation side-effects of conformal and conventional radiotherapy in prostate cancer: a randomised trial. Lancet, 353, 267–272.CrossRefGoogle ScholarPubMed
Fraass, B., Doppke, K., Hunt, M., et al. (1998). American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. Med. Phys., 25, 1773–1829.CrossRefGoogle Scholar
Gregoire, V., Coche, E., Cosnard, G., et al. (2000). Selection and delineation of lymph node target volumes in head and neck conformal radiotherapy. Proposal for standardizing terminology and procedure based on the surgical experience. Radiother. Oncol., 56, 135–150.CrossRefGoogle ScholarPubMed
ICRU. (1976). ICRU Report 24. Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma rays in Radiotherapy Procedures. Bethesda, MD: International Commission on Radiation Units and Measurements.
ICRU. (1985). ICRU Report 38. Dose and Volume Specification for Reporting Intracavitary Therapy in Gynaecology. Bethesda, MD: International Commission on Radiation Units and Measurements.
ICRU. (1993). ICRU Report 50. Prescribing, Recording, and Reporting Photon Beam Therapy.Bethesda, MD: International Commission on Radiation Units and Measurements.
ICRU. (1997). ICRU Report 58. Dose and Volume Specification in Interstitial Brachytherapy. Bethesda, MD: International Commission on Radiation Units and Measurements.
ICRU. (1999). ICRU Report 62. Prescribing, Recording, and Reporting Photon Beam Therapy (Supplement to ICRU Report 50).Bethesda, MD: International Commission on Radiation Units and Measurements.
ICRU. (2004). ICRU Report 71. Prescribing, Recording, and Reporting Electron Beam Therapy. Journal of the ICRU, 4 (No. 1). Oxford: Oxford University Press.
Jansen, E. P. M., Dewit, L. G. H., van Herk, M., et al. (2000). Target volumes in radiotherapy for high-grade malignant glioma of the brain. Radiother. Oncol, 56, 151–156.CrossRefGoogle Scholar
Kehoe, T. and Rugg, L. J. (1999). From technical quality assurance of radiotherapy to a comprehensive quality of service management system. Radiother. Oncol., 51, 281–290.CrossRefGoogle ScholarPubMed
Khan, F. M. (2003). The Physics of Radiation Therapy, 3rd edn. Philadelphia, PA: Lippincott, Williams and Wilkins.Google Scholar
Klein, E. E., Harms, W. B., Low, D. A., et al. (1995). Clinical implementation of a commercial multileaf collimator: dosimetry, networking, simulation and quality assurance. Int. J. Radiat. Oncol. Biol. Phys., 33, 1195–1208.CrossRefGoogle ScholarPubMed
Lees, J., Holloway, L., Fuller, M., et al. (2005). Effect of intravenous contrast on treatment planning system dose calculations in the lung. Australas. Phys. Eng. Sci. Med., 28, 190–195.CrossRefGoogle ScholarPubMed
Logue, J. P., Sharrock, C. L., Cowan, R. A., et al. (1998). Clinical variability of target volume description in conformal radiotherapy planning. Int. J. Radiat. Oncol. Biol. Phys., 41, 929–931.CrossRefGoogle ScholarPubMed
Meredith, W. J. (1967). Radium Dosage: the Manchester System/Compiled from Articles, by Ralston Paterson (and Others). Edinburgh: Livingstone.Google Scholar
Nutting, C. M., Khoo, V. S., Walker, V., et al. (2000). A randomized study of the use of a customized immobilization system in the treatment of prostate cancer with conformal radiotherapy. Radiother. Oncol., 54, 1–9.CrossRefGoogle ScholarPubMed
Padhani, A. R., Khoo, V. S., Suckling, J., et al. (1999). Evaluating the effect of rectal distension and rectal movement on prostate gland position using cine MRI. Int. J. Radiat. Oncol. Biol. Phys., 44, 525–533.CrossRefGoogle ScholarPubMed
Pierquin, B., Dutreix, A., Paine, C. H., et al. (1978). The Paris system in interstitial radiation therapy. Acta Radiol. Oncol. Radiat. Phys. Biol., 17, 33–48.Google ScholarPubMed
Pötter, R., Haie-Meder, C., Van Limbergen, E., et al. (2006). Recommendations from gynaecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy – 3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother. Oncol., 78, 67–77.CrossRefGoogle Scholar
Purdy, J. A. (2000). Future directions in 3-D treatment planning and delivery: a physicist's perspective. Int. J. Radiat. Oncol. Biol. Phys., 46, 3–6.CrossRefGoogle ScholarPubMed
Purdy, J. A. and Harms, W. B. (1998). Quality assurance for 3D conformal radiation therapy. Strahlenther. Onkol., 174 (Suppl. 2), 2–7.Google ScholarPubMed
Senan, S., Chapet, O., Lagerwaard, F. J., et al. (2004). Defining target volumes for non-small cell lung carcinoma. Semin. Radiat. Oncol., 14, 308–314.CrossRefGoogle ScholarPubMed
Seppenwoolde, Y., Shirato, H., Kitamura, K., et al. (2002). Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int. J. Radiat. Oncol. Biol. Phys., 53, 822–834.CrossRefGoogle ScholarPubMed
Shimizu, S., Shirato, H., Kitamura, K., et al. (2000). Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers. Int. J. Radiat. Oncol. Biol. Phys., 48, 1591–1597.CrossRefGoogle ScholarPubMed
Shiu, A. S. and Mellenberg, D. E. (2001). General Practice of Radiation Oncology Physics in the 21st Century: AAPM Monograph no. 26. Madison, Wisconsin: Medical Physics Publishing.Google Scholar
The Royal College of Radiologists. (2009). Implementing Image-Guided Brachytherapy for Cervix Cancer in the UK. London: The Royal College of Radiologists.
The Royal College of Radiologists, Society and College of Radiographers, Institute of Physics and Engineering in Medicine, National Patients Safety Agency, British Institute of Radiology. (2008). Towards Safer Radiotherapy. London: The Royal College of Radiologists.
Weiss, E., and Hess, C. F. (2003). The impact of gross tumor volume (GTV) and clinical target volume (CTV) definition on the total accuracy in radiotherapy theoretical aspects and practical experiences. Strahlenther. Onkol., 179, 21–30.CrossRefGoogle ScholarPubMed
Williams, J. R. and Thwaites, D. I. (2000). Radiotherapy Physics in Practice. Oxford: Oxford University Press.Google Scholar
Bomford, C. K. and Kunkler, I. H. (2003). Walter and Miller's Textbook of Radiotherapy. Radiation Physics, Therapy and Oncology, 6th edn. Edinburgh: Churchill Livingstone.Google Scholar
Dobbs, J., Barrett, A. and Ash, D. (1999). Practical Radiotherapy Planning, 3rd edn. New York, NY: Arnold.Google Scholar
Gerbaulet, A., Pötter, R., Mazeron, J.-J., et al. (2002). The GEC ESTRO Handbook of Brachytherapy. Brussels: European Society for Therapeutic Radiology and Oncology.Google Scholar
Hoskin, P. (2006). Radiotherapy in Practice. External Beam Therapy. Oxford: Oxford University Press.Google Scholar
Hoskin, P. and Coyle, C. (2011). Radiotherapy in Practice. Brachytherapy, 2nd edn. Oxford: Oxford University Press.CrossRefGoogle Scholar
Joslin, C. A. F., Flynn, A. and Hall, E. J. (2001). Principles and Practice of Brachytherapy: Using Afterloading Systems. London: Arnold.Google Scholar

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×