Hostname: page-component-8448b6f56d-m8qmq Total loading time: 0 Render date: 2024-04-24T05:36:56.947Z Has data issue: false hasContentIssue false

Determination of analytical precision in calculating total energy expenditure using the doubly labelled water method in humans

Published online by Cambridge University Press:  05 October 2018

R. Baker
Affiliation:
Stable Isotope Facility, MRC Elsie Widdowson Laboratory, 120 Fulbourn Road, Cambridge, CB1 9NL, UK
K. Donkers
Affiliation:
Stable Isotope Facility, MRC Elsie Widdowson Laboratory, 120 Fulbourn Road, Cambridge, CB1 9NL, UK
E.R. Orford
Affiliation:
Stable Isotope Facility, MRC Elsie Widdowson Laboratory, 120 Fulbourn Road, Cambridge, CB1 9NL, UK
P.A. Singh
Affiliation:
Stable Isotope Facility, MRC Elsie Widdowson Laboratory, 120 Fulbourn Road, Cambridge, CB1 9NL, UK
M.C. Venables
Affiliation:
Stable Isotope Facility, MRC Elsie Widdowson Laboratory, 120 Fulbourn Road, Cambridge, CB1 9NL, UK
Rights & Permissions [Opens in a new window]

Abstract

Type
Abstract
Copyright
Copyright © The Authors 2018 

Doubly labelled water (DLW) is the gold standard method for estimation of free-living energy expenditure in humans(Reference Schoeller and van Santen1).  Overall precision is an important factor to consider when designing a study, as poor precision in total energy expenditure (TEE) will increase the sample size required.  The precision or analytical variation of modelled TEE is variable and dependent on the laboratory, instrumentation used and specific methodology used.  In order to calculate the precision of our methodology, we assessed the reproducibility of modelled TEE estimation over ten consecutive weeks.  Additionally, jack-knife resampling was used to assess bias within an individual estimate of TEE.

Two female participants (mean (SD): age, 40 (4) years: height, 1.66 (0.04) m: weight, 77.6 (9.1) kg) were dosed with 150 mg/kg body weight H218O (Taiyo Nippon Sanso, Tokyo, Japan) and 80 mg/kg body weight 2H2O (Cambridge Isotope Laboratories, Massachusetts, USA).  The protocol was approved by the East of England - Cambridge Central Research Ethics Committee, Cambridge, UK.   Eleven urine samples were collected, comprising of a baseline urine sample followed by ten consecutive days post-dosing, avoiding the first void of the day. To minimise freeze-thaw cycles, each urine sample was aliquoted into ten 7 ml glass bijou bottles and stored at −20 °C until analysis.  Analysis of H218O and 2H2O was conducted weekly using a continuous-flow isotope ratio mass spectrometer (AP2003, Analytical Precision Ltd., Northwich, Cheshire, UK) and a dual-inlet isotope ratio mass spectrometer (IsoPrime, GV Instruments Ltd., Wythenshawe, Manchester, UK) respectively.  Energy expenditure was modelled according to Schoeller(Reference Schoeller, Ravussin and Schutz2) and expressed in kilojoules.  Precision was calculated as the coefficient of variation.

Average daily TEE for repeat measurements was 8643 (52) kJ and 8683 (52) kJ (mean (SE)) for participant Y & W respectively.  Both had total analytical precisions of 1.20 %, with separate precisions for isotope elimination rates and dilution spaces reported in Table 1.  Jack-knife resampling confirmed a standard error of the TEE estimate of 63  and 127 kJ and a small mean bias of 12 and −15 kJ.

Table 1. Isotope elimination rates and dilution spaces for the two participants.

Preliminary data for four repeat measurements. Values are means with  %CV.

ko, oxygen elimination rate; kh, hydrogen elimination rate; No, oxygen dilution space; Nh, hydrogen dilution space; TEE, total energy expenditure.

In conclusion, the analytical precision for our laboratory when using the doubly labelled water method to calculate TEE is well within the published limits of 3-6 %(Reference Schoeller, Taylor and Shay3, Reference Cole and Coward4).

References

1.Schoeller, DA & van Santen, E. (1982) J Appl Physiol Respir Environ Exerc Physiol 53(4), 955–9.Google Scholar
2.Schoeller, DA, Ravussin, E, Schutz, Y et al. (1986) Am J Physiol 250(5 Pt 2), R823–30.Google Scholar
3.Schoeller, DA, Taylor, PB & Shay, K (1995) Obesity Research 3(1), 1520.Google Scholar
4.Cole, TJ & Coward, WA (1992) Am J Physiol 263(26), E965E973.Google Scholar
Figure 0

Table 1. Isotope elimination rates and dilution spaces for the two participants.