Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements

  • Jordan R. Moon (a1), Jeffrey R. Stout (a2), Abbie E. Smith (a2), Sarah E. Tobkin (a3), Christopher M. Lockwood (a2), Kristina L. Kendall (a2), Jennifer L. Graef (a4), David H. Fukuda (a2), Pablo B. Costa (a5), Matt S. Stock (a5), Kaelin C. Young (a6), Patrick S. Tucker (a7), Eonho Kim (a5), Trent J. Herda (a5), Ashley A. Walter (a5), Steven L. Ferguson (a8), Vanessa D. Sherk (a6) and Joel T. Cramer (a5)...

Abstract

Bioimpedance spectroscopy (BIS) has been used to track changes in total body water (TBW). Accurate TBW estimations can be influenced by both methodological and biological factors. One methodological variation that contributes to BIS TBW errors is the electrode placement. The purpose of the present study was to compare the reproducibility and validity of fixed-distance electrode placements (5 cm) with the standard single-site electrode placements. Twenty-nine subjects (fifteen men and fourteen women) participated in the reproducibility study, while sixty-nine subjects (thirty-three men and thirty-six women) participated in the validity study. The reproducibility study included two measurements that were taken 24 h apart, while the validity study consisted of a 12-week exercise intervention with measurements taken at weeks 1 and 12. TBW was estimated using BIS and 2H techniques. Reproducibility results indicated that fixed-distance electrodes reduced the day-to-day standard error of the measurement in men (from 1·13 to 0·81 litres) but not in women (0·47 litres). sem values were lower for women than for men, suggesting that BIS TBW estimates are sex dependent. Validity results produced similar accurate findings (mean difference < 0·21 litres). However, fixed-distance electrodes improved delta TBW errors (mean difference improvements>0·04 litres in men, women, and men and women combined). When tracking changes in TBW, fixed-distance electrodes may reduce reproducibility errors and allow for smaller changes to be detected. However, the reduction of reproducibility errors may be greater for men than for women. Therefore, reproducibility calculations should be based on the sex of the sample population.

  • View HTML
    • Send article to Kindle

      To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

      Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

      Find out more about the Kindle Personal Document Service.

      Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements
      Available formats
      ×

      Send article to Dropbox

      To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

      Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements
      Available formats
      ×

      Send article to Google Drive

      To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

      Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements
      Available formats
      ×

Copyright

Corresponding author

*Corresponding author: Dr J. R. Moon, fax +1 251 625 1035, email jmoon@ussa.edu

References

Hide All
1 Pace, N & Rathbun, EN (1945) Studies on body composition. III. The body water and chemically combined nitrogen content in relation to fat content. J Biol Chem 158, 685691.
2 Evans, EM, Saunders, MJ, Spano, MA, et al. (1999) Effects of diet and exercise on the density and composition of the fat-free mass in obese women. Med Sci Sports Exerc 31, 17781787.
3 Moon, JR, Eckerson, JM, Tobkin, SE, et al. (2008) Estimating body fat in NCAA Division I female athletes: a five-compartment model validation of laboratory methods. Eur J Appl Physiol 105, 119130.
4 Wang, ZM, Deurenberg, P, Guo, SS, et al. (1998) Six-compartment body composition model: inter-method comparisons of total body fat measurement. Int J Obes Relat Metab Disord 22, 329337.
5 Armstrong, LE (2005) Hydration assessment techniques. Nutr Rev 63, S40S54.
6 Dumler, F & Kilates, C (2005) Prospective nutritional surveillance using bioelectrical impedance in chronic kidney disease patients. J Ren Nutr 15, 148151.
7 Moon, JR, Tobkin, SE, Roberts, MD, et al. (2008) Total body water estimations in healthy men and women using bioimpedance spectroscopy: a deuterium oxide comparison. Nutr Metab (Lond) 5, 7.
8 Moon, JR, Smith, AE, Tobkin, SE, et al. (2009) Total body water changes after an exercise intervention tracked using bioimpedance spectroscopy: a deuterium oxide comparison. Clin Nutr 28, 516525.
9 Ward, LC, Byrne, NM, Rutter, K, et al. (1997) Reliability of multiple frequency bioelectrical impedance analysis: an intermachine comparison. Am J Hum Biol 9, 6372.
10 Ellis, KJ, Shypailo, RJ & Wong, WW (1999) Measurement of body water by multifrequency bioelectrical impedance spectroscopy in a multiethnic pediatric population. Am J Clin Nutr 70, 847853.
11 Elsen, R, SIiu, ML, Pineda, O, et al. (1987) Sources of variability in bioelectrical impedance determinations in adults. In In Vivo Body Composition, pp. 184188 [Ellis, KJ, Yasumara, S and Morgan, WD, editors]. London: Institute of Physical Sciences in Medicine.
12 Lukaski, HC (1986) Use of tetrapolar bioelectrical impedance method to assess human body composition. In Human Body Composition and Fat Patterning, pp. 143158 [Norgan, NG, editor]. Wageningen: Euronut.
13 Kaysen, GA, Zhu, F, Sarkar, S, et al. (2005) Estimation of total-body and limb muscle mass in hemodialysis patients by using multifrequency bioimpedance spectroscopy. Am J Clin Nutr 82, 988995.
14 Cornish, BH, Jacobs, A, Thomas, BJ, et al. (1999) Optimizing electrode sites for segmental bioimpedance measurements. Physiol Meas 20, 241250.
15 Lukaski, HC, Johnson, PE, Bolonchuk, WW, et al. (1985) Assessment of fat-free mass using bioelectrical impedance measurements of the human body. Am J Clin Nutr 41, 810817.
16 Kyle, UG, Bosaeus, I, De Lorenzo, AD, et al. (2004) Bioelectrical impedance analysis – part II: utilization in clinical practice. Clin Nutr 23, 14301453.
17 Bunt, JC, Lohman, TG & Boileau, RA (1989) Impact of total body water fluctuations on estimation of body fat from body density. Med Sci Sports Exerc 21, 96100.
18 Cornish, BH, Thomas, BJ & Ward, LC (1993) Improved prediction of extracellular and total body water using impedance loci generated by multiple frequency bioelectrical impedance analysis. Phys Med Biol 38, 337346.
19 Hanai, T (1968) Electrical Properties of Emulsions in Emulsion Science, Chapter 5. London/New York: Academic Press.
20 Keppel, G & Wickens, TD (2004) Design and Analysis: A Researchers Handbook, 4th ed. Upper Saddle River, NJ: Prentice Hall.
21 Weir, JP (2005) Quantifying test–retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 19, 231240.
22 Shrout, PE & Fleiss, JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86, 420428.
23 Hopkins, WG (2000) Measures of reliability in sports medicine and science. Sports Med 30, 115.
24 de Vet, HC, Terwee, CB, Knol, DL, et al. (2006) When to use agreement versus reliability measures. J Clin Epidemiol 59, 10331039.
25 Moon, JR, Smith, AE, Tobkin, SE, et al. (2009) Total body water changes after an exercise intervention tracked using bioimpedance spectroscopy: a deuterium oxide comparison. Clin Nutr 28, 516525.
26 Balady, GJ, Berra, KA, Golding, LA, et al. (2000) ACSM's Guidelines for Exercise Testing and Prescription, 6th ed. Baltimore, MD: Lippincott Williams and Wilkins.
27 Wong, WW, Lee, LS & Klein, PD (1987) Deuterium and oxygen-18 measurements on microliter samples of urine, plasma, saliva, and human milk. Am J Clin Nutr 45, 905913.
28 Schoeller, DA, van Santen, E, Peterson, DW, et al. (1980) Total body water measurement in humans with 18O and 2H labeled water. Am J Clin Nutr 33, 26862693.
29 Heyward, VH & Wagner, DR (2004) Applied Body Composition Assessments. Champaign, IL: Human Kinetics.
30 Bland, JM & Altman, DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i, 307310.
31 Evans, WD, McClagish, H & Trudgett, C (1998) Factors affecting the in vivo precision of bioelectrical impedance analysis. Appl Radiat Isot 49, 485487.

Keywords

Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements

  • Jordan R. Moon (a1), Jeffrey R. Stout (a2), Abbie E. Smith (a2), Sarah E. Tobkin (a3), Christopher M. Lockwood (a2), Kristina L. Kendall (a2), Jennifer L. Graef (a4), David H. Fukuda (a2), Pablo B. Costa (a5), Matt S. Stock (a5), Kaelin C. Young (a6), Patrick S. Tucker (a7), Eonho Kim (a5), Trent J. Herda (a5), Ashley A. Walter (a5), Steven L. Ferguson (a8), Vanessa D. Sherk (a6) and Joel T. Cramer (a5)...

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed