Case 75 - Pulmonary nodule misregistration on PET/CT

Published online by Cambridge University Press:  07 October 2011

Thomas Hartman
Affiliation:
Mayo Clinic, Rochester
Get access

Summary

Imaging description

The CT in an integrated PET/CT scanner is optimized for PET imaging. The CT is converted to provide attenuation correction and to lend structural data to the fused PET/CT image. Accurate co-registration between the CT and PET images is optimal for both attenuation correction and fusion. The CT images are generally acquired at mid-inspiration and PET images are acquired during quiet respiration. Careful patient coaching and cooperation is required to prevent a mismatch between the two datasets. Pulmonary nodules, particularly in the lung bases, have modest motion even with quiet breathing. The net result is that “hot spot” on the PET image does not match the lung nodule on the CT image. Review of the CT slices above and below the hot focus allows confident identification of this artifact (Figure 75.1). The frequency of misregistration is lessened when PET/CT is performed on scanners with six or more rows of CT detectors [1, 2].

Importance

Incorrect localization of a pulmonary FDG “hot spot” in the lungs could have several adverse consequences. At a minimum, the error is confusing, since there will be no anatomic correlate for the foci of increased FDG uptake in the lungs on CT. Worse, a patient may be incorrectly assumed to have low FDG uptake in the pulmonary nodule so that the nodule is not reported as having high malignant potential.

Type
Chapter
Information
Pearls and Pitfalls in Thoracic Imaging
Variants and Other Difficult Diagnoses
, pp. 196 - 197
Publisher: Cambridge University Press
Print publication year: 2011

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Bockisch, ABeyer, TAntoch, GPositron emission tomography/computed tomography–imaging protocols, artifacts and pitfallsMol Imaging Biol 2004 6 188CrossRefGoogle ScholarPubMed
Beyer, TRosenbaum, SVeit, PRespiration artifacts in whole-body (18)F-FDG PET/CT studies with combined PET/CT tomographs employing spiral CT technology with 1 to 16 detector rowsEur J Nucl Med Mol Imaging 2005 32 1429CrossRefGoogle ScholarPubMed

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×