CHAPTER 10 - Procedure for Measuring Integrated Intensities

Published online by Cambridge University Press:  21 May 2010

Get access

Summary

In this chapter we shall describe the main features in the procedure for measuring the set of relative integrated intensities ρhkl of the crystal. We shall assume that the difTractometer has been correctly positioned and alined with respect to the incident beam. The methods of making these adjustments are fully described by Furnas (1957).

Choice of crystal size and shape

To minimize errors arising from the effects of absorption, extinction and simultaneous reflexions, the crystal must be as small as possible. The ultimate limit to the specimen size is determined by counting statistics (see §10.7), and, because of the greater intensity of primary X–ray beams compared with monochromatic neutron beams, the crystal size tends to be much larger in neutron diffraction. In the early period of neutron diffraction the linear dimensions of the crystal were measured in centimetres, but now, with the availability of high–flux reactors and with the use of automatic methods to speed up the collection of data, the size of crystal is down to 1 or 2 mm.

On p. 240 we gave the criterion μR ∼ 1 as a rough indication of the optimum size of the crystal, where μ is the linear absorption coefficient and R the average crystal radius. From Table XIX, this corresponds to an average radius for X–ray work of about 0.1 mm. The diffracted intensity decreases with the size of the unit cell, so that larger specimens are used when the cell size is large and the absorption is relatively low.

Type
Chapter
Information
Publisher: Cambridge University Press
Print publication year: 1966

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×