Signal quality indices for state space electrophysiological signal processing and vice versa

from Part II - State space methods for clinical data

Published online by Cambridge University Press:  05 October 2015

J. Oster
Affiliation:
University of Oxford
G. D. Clifford
Affiliation:
University of Oxford
Zhe Chen
Affiliation:
New York University
Get access

Summary

Background

Traditional approaches to physiological signal processing have focused on highly sensitive, and less specific detection techniques, generally with the expectation that an expert will overread the results and deal with the false positives. However, acquisition of physiological signals has become increasingly routine in recent years, and clinicians are often fed large flows of data which can become rapidly unmanageable and lead to missed important events and alarm fatigue (Aboukhalil et al. 2008).

Ignoring the nascent world of the quantified self for now, which itself has the potential to swamp medical practitioners with all manner of noise, there are two obvious examples of this paradigm. First, each intensive care unit (ICU) generates an enormous quantity of physiological data, up to 1GB per person per day (Clifford et al. 2009) More than 5 million patients are admitted annually to ICUs in the United States, with Europe and the rest of the world, rapidly catching up (Mullins et al. 2013; Rhodes et al. 2012). This can be attributed in part to the aging global population. ICU patients are a heterogeneous population, but all require a high level of acute care, with numerous bedside monitors. Patients in the ICU often require mechanical ventilation or cardiovascular support and invasive monitoring modalities and treatments (e.g., hemodialysis, plasmapheresis and extracorporeal membrane oxygenation). With an ever increasing reliance on technology to keep critically ill patients alive, the number of ICU beds in the US has grown significantly, to an estimated 6000 or more (Rhodes et al. 2012). Assuming an average ICU bed occupancy of 68.2% (Wunsch et al. 2013), the sum total of all bedside data generated in the US is over a petabyte of data each year. Multi-terabyte ICU databases are therefore becoming available (Saeed et al. 2011), and include parameters such as the ECG, the photoplethysmogram, arterial blood pressure and respiratory effort.

Type
Chapter
Information
Publisher: Cambridge University Press
Print publication year: 2015

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Aboukhalil, A., Nielsen, L., Saeed, M., Mark, R.G. & Clifford, G.D. (2008). Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform. Journal of Biomedical Informatics 41(3), 442–451.Google Scholar
Allen, J. & Murray, A. (1996). Assessing ECG signal quality on a coronary care unit. Physiological Measurement 17(4), 249–258.Google Scholar
Behar, J., Johnson, A. E. W., Oster, J. & Clifford, G.D. (2013b). An echo state neural network for foetal ECG extraction optimized by random search. In NIPS Workshop on Machine Learning for Clinical Data Analysis and Healthcare, Lake Tahoe, USA.Google Scholar
Behar, J., Oster, J., Li, Q. & Clifford, G.D. (2013a). ECG signal quality during arrhythmia and its application to false alarm reduction. IEEE Transactions on Biomedical Engineering 60, 1660–1666.Google Scholar
Bergstra, J. & Bengio, Y. (2012). Random search for hyper-parameter optimization. Journal of Machine Learning Research 13, 281–305.Google Scholar
Chambrin, M. -C., Ravaux, P., Calvelo-Aros, D., Jaborska, A., Chopin, C. & Boniface, B. (1999). Multicentric study of monitoring alarms in the adult intensive care unit (ICU): a descriptive analysis. Intensive Care Medicine 25(12), 1360–1366.Google Scholar
Chambrin, M. -C. et al. (2001). Alarms in the intensive care unit: how can the number of false alarms be reduced?Critical Care 5(4), 184–188.Google Scholar
Chen, L., McKenna, T., Reisner, A. & Reifman, J. (2006). Algorithms to qualify respiratory data collected during the transport of trauma patients. Physiological Measurement 27(9), 797.Google Scholar
Clifford, G. D., Azuaje, F. & McSharry, P. (2006). Advanced Methods and Tools for ECG Data Analysis, Artech House.
Clifford, G. D., Behar, J., Li, Q. & Rezek, I. (2012). Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms. Physiological Measurement 33(9), 1419–1433.Google Scholar
Clifford, G. D. & Clifton, D. (2012). Wireless technology in disease management and medicine. Annual Review of Medicine 63, 479–492.Google Scholar
Clifford, G. D. & Moody, G. B. (2012). Signal quality in cardiorespiratory monitoring. Physiological Measurement 33(9).Google Scholar
Clifford, G. D., Long, W. J., Moody, G. B. & Szolovits, P. (2009). Robust parameter extraction for decision support using multimodal intensive care data. Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences 367(1887), 411–429.Google Scholar
Ebrahim, M. H., Feldman, J. M. & Bar-Kana, I. (1997). A robust sensor fusion method for heart rate estimation. Journal of Clinical Monitoring 13(6), 385–393.Google Scholar
Feldman, J. M., Ebrahim, M. H. & Bar-Kana, I. (1997). Robust sensor fusion improves heart rate estimation: Clinical evaluation. Journal of Clinical Monitoring 13(6), 379–384.Google Scholar
Friesen, G. M., Jannett, T. C., Jadallah, M. A., Yates, S. L., Quint, S. R. & Nagle, H. T. (1990). A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Transactions on Biomedical Engineering 37(1), 85–98.Google Scholar
Goldberger, A. L., Amaral, L. A., Glass, L., Hausdorff, J. M., Ivanov, P. C., Mark, R. G., Mietus, J. E., Moody, G. B., Peng, C. -K. & Stanley, H. E. (2000). Physiobank, Physiotoolkit, and Physionet components of a new research resource for complex physiologic signals. Circulation 101(23), e215–e220.Google Scholar
Hamilton, P. S. & Tompkins, W. J. (1986). Quantitative investigation of QRS detection rules using the MIT/BIH arrhythmia database. IEEE Transactions on Biomedical Engineering 33(12), 1157–1165.Google Scholar
Hug, C. W., Clifford, G. D. & Reisner, A. T. (2011). Clinician blood pressure documentation of stable intensive care patients: an intelligent archiving agent has a higher association with future hypotension. Critical Care Medicine 39(5), 1006–1014.Google Scholar
Jakob, S., Korhonen, I., Ruokonen, E., Virtanen, T., Kogan, A. & Takala, J. (2000). Detection of artifacts in monitored trends in intensive care. Computer Methods and Programs in Biomedicine 63(3), 203–209.Google Scholar
Johnson, A. E., Cholleti, S. R., Buchman, T. G. & Clifford, G. D. (2013). Improved respiration rate estimation using a Kalman filter and wavelet cross-coherence. In Proceedings of Computing in Cardiology, pp. 791–794.Google Scholar
Kaiser, W. & Findeis, M. (2000). Novel signal processing methods for exercise ECG. International Journal on Bioelectromagnetism 2, 1–4.Google Scholar
Kalkstein, N., Kinar, Y., Na'aman, M., Neumark, N. & Akiva, P. (2011). Using machine learning to detect problems in ECG data collection. In Proceedings of Computing in Cardiology, pp. 437–440.Google Scholar
Kohler, B. -U., Hennig, C. & Orglmeister, R. (2002). The principles of software QRS detection. IEEE Magazine on Engineering in Medicine and Biology 21(1), 42–57.Google Scholar
Langley, P., Di Marco, L. Y., King, S., Duncan, D., Di Maria, C., Duan, W., Bojarnejad, M., Zheng, D., Allen, J. & Murray, A. (2011). An algorithm for assessment of quality of ECGs acquired via mobile telephones. In Proceedings of Computing in Cardiology, 281–284.Google Scholar
Lawless, S. T. (1994). Crying wolf: false alarms in a pediatric intensive care unit. Critical Care Medicine 22(6), 981–985.Google Scholar
Li, Q. & Clifford, G. D. (2012). Signal quality and data fusion for false alarm reduction in the intensive care unit. Journal of Electrocardiology 45(6), 596–603.Google Scholar
Li, Q., Mark, R. G. & Clifford, G. D. (2008). Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filter. Physiological Measurement 29(1), 15–32.Google Scholar
Llamedo, M. & Martínez, J. P. (2011). Heartbeat classification using feature selection driven by database generalization criteria. IEEE Transactions on Biomedical Engineering 58(3), 616–625.Google Scholar
Markou, M. & Singh, S. (2003). Novelty detection: a review – part 1: statistical approaches. Signal Processing 83(12), 2481–2497.Google Scholar
McSharry, P. E., Clifford, G. D., Tarassenko, L. & Smith, L. A. (2003). A dynamical model for generating synthetic electrocardiogram signals. IEEE Transactions on Biomedical Engineering 50(3), 289–294.Google Scholar
Moody, B. E. (2011). Rule-based methods for ECG quality control. In Proceedings of Computing in Cardiology, pp. 361–363.Google Scholar
Moody, G. B. & Mark, R. G. (1989). QRS morphology representation and noise estimation using the Karhunen–Loeve transform. In Proceedings of Computers in Cardiology, pp. 269–272.Google Scholar
Moody, G. B. & Mark, R. G. (2001). The impact of the MIT-BIH arrhythmia database. IEEE Magazine on Engineering in Medicine and Biology 20(3), 45–50.Google Scholar
Mullins, P. M., Goyal, M. & Pines, J. M. (2013). National growth in intensive care unit admissions from emergency departments in the United States from 2002 to 2009. Academic Emergency Medicine 20(5), 479–486.Google Scholar
Murphy, K. (1998). Switching Kalman filters. Technical report.
Nemati, S., Malhotra, A. & Clifford, G. D. (2010). Data fusion for improved respiration rate estimation. EURASIP Journal on Advances in Signal Processing 2010.Google Scholar
Oster, J., Behar, J., Johnson, A. E. W., Sayadi, O., Nemati, S. & Clifford, G. D. (2015). Semisupervised ECG ventricular beat classification with novelty detection based on switching Kalman filters. IEEE Transactions on Biomedical Engineering, in press.Google Scholar
Pan, J. & Tompkins, W. J. (1985). A real-time QRS detection algorithm. IEEE Transactions on Biomedical Engineering0 (3), 230–236.Google Scholar
Pimentel, M. A., Clifton, D. A., Clifton, L. & Tarassenko, L. (2014). A review of novelty detection. Signal Processing 99, 215–249.Google Scholar
Price, J. D., Tarassenko, L. & Townsend, N. W. (2010), Combining measurements from different sensors. US Patent 7,647,185.
Quinn, J., Williams, C. & McIntosh, N. (2009). Factorial switching linear dynamical systems applied to physiological condition monitoring. IEEE Transactions on Pattern Analysis and Machine Intelligence 31(9), 1537–1551.Google Scholar
Rhodes, A., Ferdinande, P., Flaatten, H., Guidet, B., Metnitz, P. G. & Moreno, R. P. (2012). The variability of critical care bed numbers in Europe. Intensive Care Medicine 38(10), 1647–1653.Google Scholar
Saeed, M., Villarroel, M., Reisner, A. T., Clifford, G. D., Lehman, L. -W., Moody, G. B., Heldt, T., Kyaw, T. H., Moody, B. & Mark, R. G. (2011). Multiparameter intelligent monitoring in intensive care II (MIMIC-II): a public-access intensive care unit database. Critical Care Medicine 39(5), 952–960.Google Scholar
Sameni, R., Shamsollahi, M. B., Jutten, C. & Clifford, G. D. (2007). A nonlinear bayesian filtering framework for ECG denoising. IEEE Transactions on Biomedical Engineering 54(12), 2172–2185.Google Scholar
Sayadi, O., Shamsollahi, M. B. & Clifford, G. D. (2010). Robust detection of premature ventricular contractions using a wave-based Bayesian framework. IEEE Transactions on Biomedical Engineering 57(2), 353–362.Google Scholar
Silva, I., Moody, G. B. & Celi, L. (2011). Improving the quality of ECGs collected using mobile phones: the Physionet/Computing in Cardiology Challenge 2011. In Proceedings of Computing in Cardiology, pp. 273–276.Google Scholar
Sittig, D. F. & Factor, M. (1990). Physiologic trend detection and artifact rejection: a parallel implementation of a multi-state Kalman filtering algorithm. Computer Methods and Programs in Biomedicine 31(1), 1–10.Google Scholar
Tarassenko, L., Townsend, N., Clifford, G., Mason, L., Burton, J. & Price, J. (2001). Medical signal processing using the software monitor. In A DERA/IEE Workshop on Intelligent Sensor Processing (Ref. No. 2001/050), IET, pp. 3/1–3/4.Google Scholar
Tat, T. H. C., Xiang, C. & Thiam, L. E. (2011). Physionet challenge 2011: improving the quality of electrocardiography data collected using real time QRS-complex and T-wave detection. In Proceedings of Computing in Cardiology, pp. 441–444.Google Scholar
Tsanas, A., Zañartu, M., Little, M. A., Fox, C., Ramig, L. O. & Clifford, G. D. (2014). Robust fundamental frequency estimation in sustained vowels: detailed algorithmic comparisons and information fusion with adaptive Kalman filtering. Journal of Acoustical Society of America 135(5), 2885–2901.Google Scholar
Tsien, C. L. & Fackler, J. C. (1997). Poor prognosis for existing monitors in the intensive care unit. Critical Care Medicine 25(4), 614–619.Google Scholar
Tsien, C. L., Kohane, I. S. & McIntosh, N. (2001). Building ICU artifact detection models with more data in less time. In Proceedings of the AMIA Symposium, pp. 706–710.Google Scholar
Wartzek, T., Brueser, C., Walter, M. & Leonhardt, S. (2013). Robust sensor fusion of unobtrusively measured heart rate. IEEE Journal of Biomedical and Health Informatics 18, 654–660.Google Scholar
Wunsch, H., Wagner, J., Herlim, M., Chong, D. H., Kramer, A. A. & Halpern, S. D. (2013). ICU occupancy and mechanical ventilator use in the United States. Critical Care Medicine 41(12), 2712–2719.Google Scholar
Xia, H., Garcia, G. A., McBride, J. C., Sullivan, A., De Bock, T., Bains, J., Wortham, D. C. & Zhao, X. (2011). Computer algorithms for evaluating the quality of ECGs in real time. In Proceedings of Computing in Cardiology, pp. 369–372.Google Scholar
Yang, P., Dumont, G. A. & Ansermino, J. M. (2009). Sensor fusion using a hybrid median filter for artifact removal in intraoperative heart rate monitoring. Journal of Clinical Monitoring and Computing 23(2), 75–83.Google Scholar
Zong, W., Moody, G. & Jiang, D. (2003). A robust open-source algorithm to detect onset and duration of QRS complexes. In Proceedings of Computers in Cardiology, pp. 737–740.Google Scholar
Zong, W., Moody, G. & Mark, R. (2004). Reduction of false arterial blood pressure alarms using signal quality assessment and relationships between the electrocardiogram and arterial blood pressure. Medical and Biological Engineering and Computing 42(5), 698–706.Google Scholar

Save book to Kindle

To save this book to your Kindle, first ensure coreplatform@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about saving to your Kindle.

Note you can select to save to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Available formats
×

Save book to Dropbox

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Dropbox.

Available formats
×

Save book to Google Drive

To save content items to your account, please confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account. Find out more about saving content to Google Drive.

Available formats
×