Performance of Portable Ventilators for Mass-Casualty Care

  • Thomas C. Blakeman (a1), Dario Rodriquez (a2), Warren C. Dorlac (a2), Dennis J. Hanseman (a1) (a2), Ellie Hattery (a3) and Richard D. Branson (a1)...

Abstract

Introduction: Disasters and mass-casualty scenarios may overwhelm medical resources regardless of the level of preparation. Disaster response requires medical equipment, such as ventilators, that can be operated under adverse circumstances and should be able to provide respiratory support for a variety of patient populations.

Objective: The objective of this study was to evaluate the performance of three portable ventilators designed to provide ventilatory support outside the hospital setting and in mass-casualty incidents, and their adherence to the Task Force for Mass Critical Care recommendations for mass-casualty care ventilators.

Methods: Each device was evaluated at minimum and maximum respiratory rate and tidal volume settings to determine the accuracy of set versus delivered VT at lung compliance settings of 0.02, 0.08 and 0.1 L/cm H20 with corresponding resistance settings of 10, 25, and 5 cm H2O/L/sec, to simulate patients with ARDS, severe asthma, and normal lungs. Additionally, different FIO2 settings with each device (if applicable) were evaluated to determine accuracy of FIO2 delivery and evaluate the effect on delivered VT. Ventilators also were tested for duration of battery life.

Results: VT decreased with all three devices as compliance decreased. The decrease was more pronounced when the internal compressor was activated. At the 0.65 FIO2 setting on the MCV 200, the measured FIO2 varied widely depending on the set VT. Battery life range was 311-582 minutes with the 73X having the longest battery life. Delivered VT decreased toward the end of battery life with the SAVe having the largest decrease. The respiratory rate on the SAVe also decreased approaching the end of battery life.

Conclusion: The 73X and MCV 200 were the closest to satisfying the Task Force for Mass Critical Care requirements for mass casualty ventilators, although neither had the capability to provide PEEP. The 73X provided the most consistent tidal volume delivery across all compliances, had the longest battery duration and the least decline in VT at the end of battery life.

Copyright

Corresponding author

Correspondence: Thomas C. Blakeman University of CincinnatiDepartment of SurgeryDivision of Trauma and Critical Care231 Albert Sabin WayCincinnati, Ohio 45267-0558 USA E-mail: thomas.blakeman@hotmail.com

References

Hide All
Wilgis, J: Strategies for providing mechanical ventilation in mass casualty incident: Distribution versus stockpiling. Respiratory Care 2008;53(1):96103.
ECDC: Influenza A(H1N1) Risk Assessment, December 17, 2009. available at: http://ecdc.europa.eu/en/files/pdf/Health_topics/090720_Influenza_A%28HINI%29_Risk_Assessment.pdf. Accessed on 29 June 2010.
Gray, B, Herbert, K: Hospitals in Hurricane Katrina: Challenges facing custodial institutions in a disaster. Available at http://www.urban.org/url.cfm?ID=411348. Accessed October 4, 2010.
FEMA: Texas agencies continue assistance to areas flooded by tropical storm. Available at http://www.fema.gov/news/newsrelease.fema?id=6316. Accessed October 4, 2010.
American Association for Respiratory Care: Guidelines for Acquisition of Ventilators to Meet Demands for Pandemic Flu and Mass Casualty Incidents, American Association for Respiratory Care. Available at http://www.aarc.org/resources/vent_guidelines_08.pdf. Accessed October 4, 2010.
Branson, RD, Johannigman, JA, Daugherty, EL, Rubinson, L: Surge capacity mechanical ventilation. Respiratory Care 2008;53(1):7890.
Rubinson, L, Hick, JL, Curtis, R, et al: Definitive care for the critically ill during a disaster: Medical resources for surge capacity. Chest 2008;133:32S50S.
Mass Casualty Ventilator Models: MCV 200 & MCV 200 B Operator’s Manual. Allied Healthcare Products Inc., St. Louis, MO.
SAVE operator’s manual. AutoMedx Inc., Germantown, MD.
Blakeman, T, Rodriquez, D, Petro, M, Dorlac, W, Branson, R: Laboratory evaluation of the SAVe simplified automated ventilator. Mil Med (in press).
Campbell, R, Johannigman, J, Branson, R, Austin, P, Matacia, G, Banks, G: Battery duration of portable ventilators: Effects of control variable, positive end-expiratory pressure, and inspired oxygen concentration. Respiratory Care 47(10):11731183.
Uni-Vent Model 73X Operator’s Manual. Impact Instrumentation Inc., West Caldwell, NJ.

Keywords

Performance of Portable Ventilators for Mass-Casualty Care

  • Thomas C. Blakeman (a1), Dario Rodriquez (a2), Warren C. Dorlac (a2), Dennis J. Hanseman (a1) (a2), Ellie Hattery (a3) and Richard D. Branson (a1)...

Metrics

Altmetric attention score

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed