Scintillator - Photodiode Linear Arrays for X-Ray Inspection System

  • S. M. Ignatov (a1), V.N. Potapov (a1), A. V. Fedin (a1), V. M. Chirkin (a1), L. I. Urutskoev (a1), V. V. Gostilo (a2), V. V. Kondrashov (a2) and A. D. Sokolov (a2)...

Abstract

The evaluations of the following performances of detectors for design of the radiographic system were carried out by mathematical simulation method and experimentally: sensitivity to absorbed energy, noise level, form of apparatus function for spatial resolution of detectors array.

The results of calculations were used for designing of CdWO4-Si arrays with optimal sizes of scintillator and photodiode for resolution ability. At the given stage the quality of radiation detectors was evaluated by spectroscopic methods. At the bias voltage on photodiode 24 V, the sensitivity of detectors has made up (0,95–1,00) × 10−21 C / eV, noise level - (1,50–1,75) ×10−16 C together with electronic noise. Dynamic range at absorbed energy by separate detector is more than 105, what is sufficient to get required image contrast.

Copyright

References

Hide All
[1]. METORSCAN. Leaflet of METOREX International Oy, 1997.

Scintillator - Photodiode Linear Arrays for X-Ray Inspection System

  • S. M. Ignatov (a1), V.N. Potapov (a1), A. V. Fedin (a1), V. M. Chirkin (a1), L. I. Urutskoev (a1), V. V. Gostilo (a2), V. V. Kondrashov (a2) and A. D. Sokolov (a2)...

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed