Total scatter factor for small fields in radiotherapy: a dosimetric comparison

  • Qurat-ul-ain Shamsi (a1), Saeed Ahmad Buzdar (a1), Saima Altaf (a1), Atia Atiq (a1), Maria Atiq (a1) and Khalid Iqbal (a2)...

Abstract

Purpose

Small field dosimetry is complicated and accuracy in the measurement of total scatter factor (TSF) is crucial for dosimetric calculations, in making optimum intensity-modulated radiotherapy plans for treating small target volumes. In this study, we intended to determine the TSF measuring properties of CC01 and CC04 detectors for field sizes ranging from sub-centimetre to the centimetre fields.

Material and methods

CC01 and CC04 chamber detectors were used to measure TSF for 6 and 18 MV photon beam delivered from the linear accelerator, through small fields in a water phantom. Small fields were created by collimator jaws and multi-leaf collimators separately, with field sizes ranging from 0·6 to 10 cm2 and 0·5 to 20 cm2, respectively.

Results

CC01 measured TSF at all the given field sizes created by jaws and multi-leaf collimators for both 6 and 18 MV beams whereas CC04 could not measure TSF for field sizes <1 cm2 due to volume averaging and perturbation effects.

Conclusion

CC01 was shown to be effective for measurement of TSF in sub-centimetre field sizes. CC01 can be employed to measure other dosimetric quantities in small fields using different energy beams.

Copyright

Corresponding author

Correspondence to: Qurat-ul-ain Shamsi, Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab 63100, Pakistan. E-mail: annieshamsi@hotmail.com. Tel: +92622875063.

References

Hide All
1. Chaudhari, S H, Dobhal, R, Kinhikar, R A, Kadam, S S, Deshpande, D D. Measurement of total scatter factor for stereotactic cones with plastic scintillation detector. J Med Phys 2017; 42 (1): 913.
2. Sharma, S D. Challenges of small photon field dosimetry are still challenging. J Med Phys 2014; 39 (3): 131.
3. Das, I J, Ding, G X, Ahnesjö, A. Small fields: nonequilibrium radiation dosimetry. Med Phys 2008; 35: 206215.
4. Aspradakis, M, Byrne, J, Palmans, H et al Small field MV photon dosimetry. IPEM Report No. 103. York, UK: Institute of Physics and Engineering in Medicine, 2010.
5. Das, IJ, Morales, J, Francescon, P. Small field dosimetry: What have we learnt? In Massillon-JL G, Fossion R, Mendez IM, Avila-Rodriguez MA, López-Pérez DO (eds). AIP Conference Proceedings, Volume 1747, No. 1. Indianapolis, IN, USA: AIP Publishing, 2016: 060001.
6. Francescon, P, Cora, S, Cavedon, C, Scalchi, P, Stancanello, J. CyberKnife dosimetric beam characteristics: comparison between experimental results and Monte Carlo simulation Robot. Radiosurgery 2005; 1 (1): 7180.
7. Francescon, P, Kilby, W, Satariano, N. Monte Carlo simulated correction factors for output factor measurement with the CyberKnife system? Results for new detectors and correction factor dependence on measurement distance and detector orientation. Phys Med Biol 2014; 59: N11N17.
8. Alfonso, R, Andreo, P, Capote, R et al. A new formalism for reference dosimetry of small and nonstandard fields. Med Phys 2008; 35 (11): 51795186.
9. Cranmer-Sargison, G, Weston, S, Evans, J A, Sidhu, N P, Thwaites, D I. Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors. Med Phys 2011; 38: 65926602.
10. Azimi, R, Alaei, P, Higgins, P. The effect of small field output factor measurements on IMRT dosimetry. Med Phys 2012; 39 (8): 46914694.
11. Birgani, M J, Chegeni, H, Behrooz, M A, Bagheri, M, Danyaei, A, Shamsi, A. An analytical method to calculate phantom scatter factor for photon beam accelerators. Electron Physician 9 (1): 35233528.
12. Nasir, M K, Amjad, N, Razzaq, A, Siddique, T. Measurement and analysis of PDDs profile and output factors for small field sizes by CC13 and micro-chamber CC01. Int J Med Phys Clin Eng Radiat Oncol 2017; 6 (01): 36.
13. Podgorsak, E B. Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency, 2005.
14. Khan, F M, John, P G. Khan’s the Physics of Radiation Therapy. Philadelphia, PA: Lippincott Williams & Wilkins, 2014.
15. Arnfield, M R, Siebers, J V, Kim, J O, Wu, Q, Keall, P J, Mohan, R. A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy. Med Phys 2000; 27 (10): 22312241.
16. Jabbari, K, Akbari, M, Tavakoli, M B, Amouheidari, A. Dosimetry and evaluating the effect of treatment parameters on the leakage of multi leaf collimators in ONCOR linear accelerators. Adv Biomed Res 2016; 5 (1): 193200.
17. Würfel, J U. Dose measurements in small fields. Med Phys 2013; 1 (1): 8190.
18. Godson, H F, Ravikumar, M, Sathiyan, S, Ganesh, K M, Ponmalar, Y R, Varatharaj, C. Analysis of small field percent depth dose and profiles: comparison of measurements with various detectors and effects of detector orientation with different jaw settings. J Med Phys 2016; 41 (1): 12.
19. Apipunyasopon, L, Srisatit, S, Phaisangittisakul, N. An investigation of the depth dose in the build-up region, and surface dose for a 6-MV therapeutic photon beam: Monte Carlo simulation and measurements. J Radiat Res 2013; 54: 374382.
20. Tyler, M K, Liu, P Z, Lee, C, McKenzie, D R, Suchowerska, N. Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators. J Appl Clin Med Phys 2016; 17 (3): 223235.
21. Azangwe, G, Grochowska, P, Georg, D et al. Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams. Med Phys 2014; 41 (7): 072103.
22. Reduan, Aa, Mazurawati, Mb, Nur Iziana, Ma, Nik Ruzman, NIb, Ahmad, Za, Ahmad Lutfi, Yb. Verification of relative output factor (ROF) measurement for radiosurgery small photon beams. Health Environ J 2016; 7 (2): 2031.
23. Washington, C M, Dennis, T. Leaver. Principles and Practice of Radiation Therapy. Amsterdam: Elsevier Health Sciences, 2015: 483484.
24. Arnfield, M R, Siebers, J V, Kim, J O, Wu, Q, Keall, P J, Mohan, R. A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy. Med Phys 2000; 27 (10): 22312241.

Keywords

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed