Sequential cohort study comparing chlorine dioxide wipes with automated washing for decontamination of flexible nasendoscopes

  • C Q Phua (a1), Y Mahalingappa (a1) and Y Karagama (a1)

Abstract

Introduction:

Flexible nasoendoscope is an important tool in otorhinolaryngology practice. The endoscope needs to be decontaminated prior to use in the next patient. The 2005 ENT-UK guidance for cleaning fibre-optic laryngoscopes stated that the ideal disinfecting agent and process should be effective and have low capital and maintenance costs.

Objective:

To compare the efficacy and cost-effectiveness of chlorine dioxide wipes versus automated washer, for decontamination of flexible nasendoscopes.

Methods:

A sequential cohort, in vitro study was performed to test the efficacy of chlorine dioxide wipes and automated washer. Costs were also calculated.

Results:

After deliberate bacterial contamination of the nasendoscope and subsequent decontamination, swab samples from the endoscope showed Staphylococcus epidermidis growth in 2 per cent (1/50 swabs) of the chlorine dioxide wipe group and in 28 per cent (14/50 swabs) of the automated washer group (p = 0.00). Based on a projected 10-year cost calculation, the automated washer was cheaper.

Conclusion:

Further studies are required to test whether these results are replicable. A similar study should be performed using real patients, to check the significance of improper decontamination.

Copyright

Corresponding author

Address for correspondence: Mr Yakubu Karagama, Department of ENT Surgery, Tameside General Hospital, Fountain Street, Ashton OL6 9RW, UK E-mail: y.karagama@nhs.net

Footnotes

Hide All

Presented as a poster at the North of England Otolaryngology Spring Meeting, 11 March 2011, Darlington, UK

Footnotes

References

Hide All
1 Foweraker, JE. The laryngoscope as a potential source of crossinfection. J Hosp Infect 1995;29:315–6
2 Esler, MD, Baines, LC, Wilkinson, DJ, Langford, RM. Decontamination of laryngoscopes: a survey of national practice. Anaesthesia 1999;54:587–92
3 Lim, M, Gupta, D. Flexible naso-endoscopic decontamination – rationalizing the next step forward. J Hosp Infect 2006;62:136–40
4 Lubbe, DE, Fagan, JJ. South African survey on disinfection techniques for the flexible nasopharyngoscope. J Laryngol Otol 2003;117:811–4
5 Banfield, GK, Hinton, AE. A national survey of disinfection techniques for flexible nasoendoscopes in UK ENT out-patient departments. J Laryngol Otol 2000;113:202–4
6 Guidance of decontamination and sterilization of rigid and flexible endoscopes. In: http://www.entuk.org/news/index_html?news_id=284 [6 February 2011]
7 Guidelines for cleaning fibreoptic laryngoscopes. In: www.entuk.org/members/publications/scopesfullweb2.pdf [6 February 2011]
8 Rejchrt, S, Cermal, P, Pavlatova, L, Mickova, E, Bures, J. Bacteriologic testing of endoscopes after high-level disinfection. Gastrointest Endosc 2004;60:68
9 Muscarella, LF. Prevention of disease transmission during flexible laryngoscopy. Am J Infect Control 2007;35:536–44
10 Birnie, GG, Quigley, EM, Clements, GB, Follet, EAC, Watkinson, G. Endoscopic transmission of hepatitis B virus. Gut 1983;24:171–4
11 Anonymous. Colonoscopes may spread HCV and HPV. AIDS Patient Care STDS 2003;7:257–8
12 Michele, TM, Cronin, WA, Graham, NM, Dwyer, DM, Pope, DS, Harrington, S et al. Transmission of Mycobacterium tuberculosis by a fiberoptic bronchoscope. Identification by DNA fingerprinting. JAMA 1997;278:1093–5
13 Kaczmarek, RG, Moore, RM Jr, McCrohan, J, Goldmann, DA, Reynolds, C, Caquelin, C et al. Multi-state investigation of the actual disinfection/sterilization of endoscopes in health care facilities. Am J Med 1992;92:257–61
14 Muscarella, LF. Inconsistencies in endoscope-reprocessing and infection-control guidelines: the importance of endoscope drying. Am J Gastroenterol 2006;101:2147–54
15 Nelson, DB, Jarvis, WR, Rutala, WA, Foxx-Orenstein, AE, Isenberg, G, Dash, GR et al. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes. Society for Healthcare Epidemiology of America. Infect Control Hosp Epidemiol 2003;24:532–7
16 NHS Estates/Department of Health. Health Technical Memorandum 2030. Washer – Disinfectors. London: Department of Health, 1997

Keywords

Sequential cohort study comparing chlorine dioxide wipes with automated washing for decontamination of flexible nasendoscopes

  • C Q Phua (a1), Y Mahalingappa (a1) and Y Karagama (a1)

Metrics

Altmetric attention score

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed