Hostname: page-component-848d4c4894-cjp7w Total loading time: 0 Render date: 2024-06-23T01:59:55.265Z Has data issue: false hasContentIssue false

Gram-Negative Bacteremia in Open-Heart–Surgery Patients Traced to Probable Tap-Water Contamination of Pressure-Monitoring Equipment

Published online by Cambridge University Press:  02 January 2015

Judith R. Rudnick
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
Consuelo M. Beck-Sague
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
Roger L. Anderson
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
Barbara Schable
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
J. Michael Miller
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
William R. Jarvis*
Affiliation:
Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, US Public Health Service, Department of Health and Human Services, Atlanta, Georgia
*
Hospital Infections Program, Centers for Disease Control and Prevention, 1600 Clifton Rd, Mailstop E-69, Atlanta, GA 30333

Abstract

Objective: To determine the cause(s) of an outbreak of gram-negative bacteremia (GNB) in open-heart–surgery (OHS) patients at hospital A.

Design: Case-control and cohort studies and an environmental survey.

Results: Nine patients developed GNB with Enterobacter cloacae (6), Pseudomonas aeruginosa (5), Klebsiella pneumoniae (3), Serratia marcescens (2), or Klebsiella oxytoca (1) following OHS; five of nine patients had polymicrobial bacteremia. When the GNB patients were compared with randomly selected OHS patients, having had the first procedure of the day (8 of 9 versus 12 of 27, P=.02), longer cardiopulmonary bypass (median, 122 versus 83 minutes, P=.01) or cross-clamp times (median, 75 versus 42 minutes, P=.008), intraoperative dopamine infusion (9 of 9 versus 15 of 27, P=.01), or exposure to scrub nurse 6 (6 of 9 versus 4 of 27, P=.001) were identified as risk factors. When stratified by length of the procedure, only being the first procedure of the day and exposure to scrub nurse 6 remained significant. First procedures used pressure-monitoring equipment that was assembled before surgery and left open and uncovered overnight in the operating room, whereas other procedures used pressure-monitoring equipment assembled immediately before the procedure. At night, operating rooms were cleaned by maintenance personnel who used a disinfectant-water solution sprayed through a hose connected to an automatic diluting system. Observation of the use of this hose documented that this solution could have contacted and entered uncovered pressure-monitoring equipment left in the operating room. Water samples from the hose revealed no disinfectant, but grew P aeruginosa. The outbreak was terminated by setting up pressure-monitoring equipment immediately before the procedure and discontinuing use of the hose-disinfectant system.

Conclusions: This outbreak most likely resulted from contamination of uncovered preassembled pressure-monitoring equipment by water from a malfunctioning spray disinfectant device. Pressure-monitoring equipment should be assembled immediately before use and protected from possible environmental contamination.

Type
Original Articles
Copyright
Copyright © The Society for Healthcare Epidemiology of America 1996

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Maki, DG. Epidemic nosocomial bacteremia. In: Wenzel, RP, ed. Handbook of Hospital-Acquired Infections. West Palm Beach, FL: CRC Press; 1981:371512.Google Scholar
2. Knaus, WA, Draper, EA, Wagner, DP, Zimmerman, JE. APACHE II: a severity of disease classification system. Crit Care Med 1985:13:818829.CrossRefGoogle ScholarPubMed
3. Siegel, S. Nonparametric Statistics for the Behavioral Sciences. New York, NY: McGraw-Hill; 1956:184.Google Scholar
4. Clark, WA, Hollis, DG, Weaver, RE, Riley, P. Identification of unusual pathogenic gram-negative aerobic and facultative anaerobic bacteria. Atlanta, GA: Centers for Disease Control; 1984.Google Scholar
5. Birnboim, HC, Doly, J. A rapid alkaline extraction procedure for screening recombinant plasmid DNA. Nucleic Acids Res 1979;7:15131523.CrossRefGoogle ScholarPubMed
6. Weinstein, RA, Emori, TG, Anderson, RL, Stamm, WE. Pressure transducers as a source of bacteremia after open heart surgery: report of an outbreak and guidelines for prevention. Chest 1976;69:338344.CrossRefGoogle ScholarPubMed
7. Beck-Sague, CM, Jarvis, WR. Epidemic bloodstream infections associated with pressure transducers: a persistent problem. Infect Control Hosp Epidemiol 1989;10:5459.CrossRefGoogle ScholarPubMed
8. Tenold, R, Priano, L, Kim, K, Rourke, B, Marrone, T. Infection potential of nondisposable pressure transducers prepared prior to use. Crit Care Med 1987;15:582583.CrossRefGoogle ScholarPubMed
9. Simmons, BP. Guideline for prevention of infections related to intravascular pressure-monitoring systems. Am J Infect Control 1983;11:183193.CrossRefGoogle Scholar