Hostname: page-component-8448b6f56d-m8qmq Total loading time: 0 Render date: 2024-04-25T00:16:06.120Z Has data issue: false hasContentIssue false

Epidemic Keratoconjunctivitis in a University Medical Center Ophthalmology Clinic; Need for Re-evaluation of the Design and Disinfection of Instruments

Published online by Cambridge University Press:  21 June 2016

Denise Koo
Affiliation:
School of Public Health, University of California, Berkeley, California
Brenda Bouvier
Affiliation:
Department of Infection Control, University of California, San Francisco, California
Michele Wesley
Affiliation:
School of Public Health, University of California, Berkeley, California
Paul Courtright
Affiliation:
School of Public Health, University of California, Berkeley, California
Arthur Reingold*
Affiliation:
School of Public Health, University of California, Berkeley, California
*
Earl Warren Hall, School Of Public Health, University of California, Berkeley, CA 94720

Abstract

During the period from early December 1987 to late January 1988, an outbreak of epidemic keratoconjunctivitis (EKG) caused by adenovirus type 8 occurred at a university medical center eye clinic. A telephone survey of patients revealed an attack rate of 17% among patients seen during the epidemic period. A case-control study demonstrated that development of EKC was associated with exposure to pneumotonometry and to three caregivers. After controlling for exposure to pneumo-tonometry, however, only exposure to one caregiver was significantly associated with risk of the disease. On January 19, 1988, after recognition of the outbreak, the eye clinic instituted additional infection control measures. A survey of patients seen the following week showed a substantial decrease in the attack rate (1.9%). Before these control measures were implemented, clinic personnel had followed the manufacturer's recommendations and disinfected pneumotonometer tips with 70% isopropyl alcohol, fearing that other disinfectants would corrode the instrument metal. Isopropyl alcohol, shown to have limited activity against adenovirus in vitro, also was being used to “disinfect” pneumotonometer tips between uses at six other area eye clinics polled by telephone. The results of this study demonstrate the need for changes in the design and manufacture of equipment used in the eye clinic.

Type
Original Articles
Copyright
Copyright © The Society for Healthcare Epidemiology of America 1989

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1.Barnard, DL, Dean Hart, JC, Marmion, VJ, Clarke, SKK. Outbreak in Bristol of conjunctivitis caused by adenovirus type 8 and its epidemiology and control. Br Med J. 1983;2:165169.Google Scholar
2.Dawson, CR, Hanna, L, Wood, TR, Despain, R. Adenovirus type 8 keratoconjunctivitis—epidemiologic, clinical, and microbiologic features. Am J Ophthalmol 1970;69:473480.Google Scholar
3.Ford, E, Nelson, K, Warren, D. Epidemiology of epidemic keratoconjunctivitis. Epidemiol Rev. 1987;9:244261.Google Scholar
4.Keenlyside, RA, Hierholzer, JC, D'Angelo, LJ. Keratoconjunctivitis associated with adenovirus type 37: an extended outbreak in an ophthalmologist's office. J Infect Dis. 1983;147:191198.Google Scholar
5.Buehler, JW, Finton, RJ, Goodman, RA, et al.Epidemic keratoconjunctivitis: report of an outbreak in an ophthalmology practice and recommendations for prevention. Inf Control 1984;5:390394.Google Scholar
6.D'Angelo, LJ, Hierholzer, JC, Holman, RC, Smith, JD. Epidemic keratoconjunctivitis caused by adenovirus type 8: epidemiologic and laboratory aspects of a large outbreak. Am J Epidemiol 1981;113:4449.Google Scholar
7.Centers for Disease Control. Recommendations for prevention of I IIV transmission in health care settings. MMWK. 1987;36(Suppl 2S).Google Scholar
8.Jewell, NPOn the bias of-commonly used measures of association for 2×2 tables. Biometrics. 1986;42:351358.Google Scholar
9.Nagington J., Sutehall, Whipp, P. Tonometer disinfection and viruses. Br J Ophthalmol 1983;63:674676.Google Scholar
10.Centers for Disease Control. Update: universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus, and other bloodborne pathogens in health-care settings. MMWR. 1988;37:377.Google Scholar
11.Garner, JS, Favero, MS. Guidelines for Handwashing and Hospital Environmental Control, 1985. Atlanta, Ga: Public Health Service, Centers for Disease Control;1985. Department of Health and Human Services publication 991117.Google Scholar
12.Spaulding, EH. Role of chemical disinfection in the prevention of nosocomial infections. In Proceedings of the International Conference on Nosocomial Infections. Baltimore, Md: American Hospital Association;1971:247254.Google Scholar