Extract
Scanned Probe Microscopy first received widespread recognition in the form of scanning tunneling microscopy (STM) images clearly showing atomic resolution of the Si 111 surface in the characteristic 7×7 surface reconstruction. For this sample, STM imaging under carefully controlled ultrahigh vacuum conditions reveals the clear image of each atom position within the surface unit cell with excellent contrast and clearly atomic resolution. Over the past few years, versions of scanning force microscopy (commonly referred to as atomic force microscopy or AFM) have become much more widespread than STM. A very common, and very difficult question, is: What is the resolution of AFM? The simple answer is that SPM in general, and STM and AFM in particular, routinely obtain sub-angstrom resolution—in the z axis, or the sample height direction. This high resolution capability is easily demonstrated by scanning a cleaved crystal of known lattice spacing and observing single and multiple atomic steps.
- Type
- The Limits of Image Resolution: Seeing Is Believing
- Information
- Microscopy and Microanalysis , Volume 3 , Issue S2: Proceedings: Microscopy & Microanalysis '97, Microscopy Society of America 55th Annual Meeting, Microbeam Analysis Society 31st Annual Meeting, Histochemical Society 48th Annual Meeting, Cleveland, Ohio, August 10-14, 1997 , August 1997 , pp. 1187 - 1188