Hostname: page-component-8448b6f56d-sxzjt Total loading time: 0 Render date: 2024-04-18T16:06:22.968Z Has data issue: false hasContentIssue false

Novel MEMS-Based Gas-Cell/Heating Specimen Holder Provides Advanced Imaging Capabilities for In Situ Reaction Studies

Published online by Cambridge University Press:  27 July 2012

Lawrence F. Allard*
Affiliation:
Physical Sciences Directorate, Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA
Steven H. Overbury
Affiliation:
Physical Sciences Directorate, Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA
Wilbur C. Bigelow
Affiliation:
Department of Materials Science & Engineering, University of Michigan, Ann Arbor, MI 48104, USA
Michael B. Katz
Affiliation:
Department of Materials Science & Engineering, University of Michigan, Ann Arbor, MI 48104, USA
David P. Nackashi
Affiliation:
Protochips Inc., 617 Hutton St., Suite 111, Raleigh, NC 27606, USA
John Damiano
Affiliation:
Protochips Inc., 617 Hutton St., Suite 111, Raleigh, NC 27606, USA
*
Corresponding author. E-mail: allardLFjr@ornl.gov
Get access

Abstract

In prior research, specimen holders that employ a novel MEMS-based heating technology (AduroTM) provided by Protochips Inc. (Raleigh, NC, USA) have been shown to permit sub-Ångström imaging at elevated temperatures up to 1,000°C during in situ heating experiments in modern aberration-corrected electron microscopes. The Aduro heating devices permit precise control of temperature and have the unique feature of providing both heating and cooling rates of 106°C/s. In the present work, we describe the recent development of a new specimen holder that incorporates the Aduro heating device into a “closed-cell” configuration, designed to function within the narrow (2 mm) objective lens pole piece gap of an aberration-corrected JEOL 2200FS STEM/TEM, and capable of exposing specimens to gases at pressures up to 1 atm. We show the early results of tests of this specimen holder demonstrating imaging at elevated temperatures and at pressures up to a full atmosphere, while retaining the atomic resolution performance of the microscope in high-angle annular dark-field and bright-field imaging modes.

Type
Special Section: Aberration-Corrected Electron Microscopy
Copyright
Copyright © Microscopy Society of America 2012

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

REFERENCES

Allard, L.F., Bigelow, W.C., Blom, D.A. & Liu, J. (2011). A stable double-tilt heating capability for precision atomic-level imaging of catalysts at elevated temperatures. Microsc Microanal 17(S2), 468469.CrossRefGoogle Scholar
Allard, L.F., Bigelow, W.C., Jose-Yacaman, M., Nackashi, D.P., Damiano, J. & Mick, S.E. (2009a). A new MEMS-based system for ultra-high-resolution imaging at elevated temperatures. Microsc Res Techniq 72(3), 208215.CrossRefGoogle ScholarPubMed
Allard, L.F., Borisevich, A., Deng, W., Si, R., Flytzani-Stephanopoulos, M. & Overbury, S.H. (2009b). Evolution of gold structure during thermal treatment of Au/FeO x catalysts revealed by aberration-corrected electron microscopy. J Elect Microsc 58(3), 199212.CrossRefGoogle Scholar
Boyes, E.D. & Gai, P.L. (1997). Environmental high-resolution electron microscopy and applications to chemical science. Ultramicroscopy 67(1-4), 219232.Google Scholar
Creemer, J.F., Helveg, S., Hoveling, G.H., Ullmann, S., Molenbroek, A.M., Sarro, P.M. & Zandbergen, H.W. (2008). Atomic-scale electron microscopy at ambient pressure. Ultramicroscopy 108(9), 993998.CrossRefGoogle ScholarPubMed
Gai, P.L. (1999a). Environmental high-resolution electron microscopy of gas-catalyst reactions. Top Catal 8(1-2), 97113.CrossRefGoogle Scholar
Gai, P.L. (1999b). Probing selective oxidation catalysis under reaction conditions by atomic-scale environmental high-resolution electron microscopy. Curr Opin Solid St Mater Sci 4(1), 6373.CrossRefGoogle Scholar
Gai, P.L. (2002). Developments in in situ environmental cell high-resolution electron microscopy and applications to catalysis. Top Catal 21(4), 161173.CrossRefGoogle Scholar
Gai, P.L. & Boyes, E.D. (2009). Advances in atomic resolution in situ environmental transmission electron microscopy and 1 Å aberration-corrected in situ electron microscopy. Microsc Res Techniq 72, 153164.CrossRefGoogle Scholar
Katz, M.B., Zhang, S.Y., Duan, Y.W., Wang, H.J., Fang, M.H., Zhang, K., Li, B.H., Graham, G.W. & Pan, X.Q. (forthcoming). Reversible precipitation/dissolution of precious-metal clusters in perovskite-based catalyst materials: Bulk versus surface re-dispersion. J Catal (in press).Google Scholar
Sharma, R. (2001). Design and applications of environmental cell transmission electron microscope for in situ observations of gas-solid reactions. Microsc Microanal 7(6), 494506.CrossRefGoogle Scholar
Sharma, R., Crozier, P.A., Kang, Z.C. & Eyring, L. (2004). Observation of dynamic nanostructural and nanochemical changes in ceria-based catalysts during in situ reduction. Phil Mag 84(25-26), 27312747.CrossRefGoogle Scholar
Sharma, R. & Weiss, K. (1998). Development of a TEM to study in situ structural and chemical changes at an atomic level during gas-solid interactions at elevated temperatures. Microsc Res Techniq 42(4), 270280.3.0.CO;2-U>CrossRefGoogle ScholarPubMed
van Huis, M.A., Young, N.P., Pandraud, G., Creemer, J.F., Vanmaekelbergh, D., Kirkland, A.I. & Zandbergen, H.W. (2009). Atomic imaging of phase transitions and morphology transformations in nanocrystals. Adv Mater 21, 49924995.CrossRefGoogle ScholarPubMed
Wittig, J.E., Bentley, J., Allard, L.F. & Lukehart, C.M. (2010). Evolution of FePt nanoparticle shape and L10 order during in situ annealing. Microsc Microanal 16(S2), 17541755.CrossRefGoogle Scholar
Yaguchi, T., Suzuki, M., Watabe, A., Nagakubo, Y., Ueda, K. & Kamino, T. (2011). Development of a high temperature–atmospheric pressure environmental cell for high-resolution TEM. J Elect Microsc 60(3), 217225.CrossRefGoogle ScholarPubMed