Hostname: page-component-7479d7b7d-qlrfm Total loading time: 0 Render date: 2024-07-08T18:09:57.767Z Has data issue: false hasContentIssue false

Design and Construction of a Quantitative Uniaxial Straining Stage For The Environmental SEM

Published online by Cambridge University Press:  02 July 2020

John F. Mansfield
Affiliation:
North Campus Electron Microbeam Analysis Laboratory 417 SRB, University of Michigan, 2455 Hayward, Ann Arbor, MI 48109-2143 http://emalwww.engin.umich.edu/people/jfmjfin/
Michael D. Thouless
Affiliation:
North Campus Electron Microbeam Analysis Laboratory 417 SRB, University of Michigan, 2455 Hayward, Ann Arbor, MI 48109-2143 http://emalwww.engin.umich.edu/people/jfmjfin/
Justin A. Stefano
Affiliation:
North Campus Electron Microbeam Analysis Laboratory 417 SRB, University of Michigan, 2455 Hayward, Ann Arbor, MI 48109-2143 http://emalwww.engin.umich.edu/people/jfmjfin/
Judd Holbrook
Affiliation:
North Campus Electron Microbeam Analysis Laboratory 417 SRB, University of Michigan, 2455 Hayward, Ann Arbor, MI 48109-2143 http://emalwww.engin.umich.edu/people/jfmjfin/
Get access

Extract

The environmental scanning electron microscope (or Environmental SEM) is a powerful tool in the materials science laboratory. The large sample chamber and ability to operate in a very poor vacuum (>5Torr) makes it an ideal device for a wide variety of in-situ studies. A three point bending stage has been used in the ElectroScan E3 environmental SEM at the University of Michigan for over four years. The stage is extremely flexible and can be used to view samples in both planar and transverse views. It also allows for quantitative recording of the stress and strain exerted on the sample via a load cell and a linear variable differential transducer. The output s of the load cell are acquired into an Apple Macintosh computer via a National Instruments analog-to-digital acquisition card. To extend the capability of EMAL's in-situ tensile testing laboratory, a uniaxial tensile stage has been constructed in collaboration with the Department of Mechanical Engineering and Applied Mechanics.

Type
New Trends in Scanning Electron Microscopy and Microanalysis
Copyright
Copyright © Microscopy Society of America

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1“Development of an Economical Electron Backscattering Diffraction System for an Environmental Scanning Electron Microscope” Thaveeprungsripom, V., Mansfield, J.F., Crawford, D.C. & Was, G.S., J. Mat. Res 9 (1994) pl8871894.Google Scholar
2Mansfield, John F., Thaveeprungsripom, Visit & Was, Gary S., “Design And Construction Of An Electron Back-Scattering Pattern Camera For The Environmental Scanning Electron Microscope”, Scanning 15 (1993) pIII36III37.Google Scholar
3Kalnas, C.E., Mansfield, J.F., Was, G.S., and Jones, J.W., “In-situ bend fixture for deformation and fracture studies in the environmental scanning electron microscope”, J. Vac. Sci. Technol. A 12 1994 883.CrossRefGoogle Scholar
4Mansfield, John F., Kalnas, Christine E., Wayne Jones, J., Was, Gary S. & Kalnas, Ronald S., “Design & Construction Of A Tensile Stage For In-Situ Sample Bending In The ESEM”, Scanning 15 (1993) PIII37III38.Google Scholar
7http://www.interfaceforce.com/. Interface Inc., 7401 E. Butherus Drive, Scottsdale AZ 85260 (602) 948-5555.Google Scholar
8http://www.sensotec.com/. Sensotec, 2080 Arlingate Lane. Columbus,Ohio 43228.Google Scholar
9Mansfield, John F., “Quicktime As a Storage Medium for Dynamic Data Sets in >In-Situ Electron Microscopy”, these proceedings.In-Situ+Electron+Microscopy”,+these+proceedings.>Google Scholar