Comparison of neutron contamination in small photon fields of secondary collimator jaws and circular cones

Published online by Cambridge University Press:  29 September 2023

Nooshin Banaee*
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Kiarash Goodarzi
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Elham Hosseinzadeh
Affiliation:
Department of Medical Physics, Faculty of Medicine, Jundishapur University of Medical Sciences, Ahvaz, Iran
*
Corresponding author: Dr Nooshin Banaee; Email: nooshin_banaee@yahoo.com

Abstract

Introduction:

Advanced treatment modalities involve applying small fields which might be shaped by collimators or circular cones. In these techniques, high-energy photons produce unwanted neutrons. Therefore, it is necessary to know neutron parameters in these techniques.

Materials and methods:

Different parts of Varian linac were simulated by MCNPX, and different neutron parameters were calculated. The results were then compared to photoneutron production in the same nominal fields created by circular cones.

Results:

Maximum neutron fluence for 1 × 1, 2 × 2, 3 × 3 cm2 field sizes was 165, 40.4, 19.78 (cm–2.Gy-1 × 106), respectively. The maximum values of neutron equivalent doses were 17.1, 4.65, 2.44 (mSv/Gy of photon dose) for 1 × 1, 2 × 2, 3 × 3 cm2 field size, respectively, and maximum neutron absorbed doses reached 903, 253, 131 (µGy/Gy photon dose) for 1 × 1, 2 × 2, 3 × 3 cm2 field sizes, respectively.

Conclusion:

Comparing the results with those in the presence of circular cones showed that circular cones reduce photoneutron production for the same nominal field sizes.

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Hosseinzadeh, E, Banaee, N, Nedaie, HA. Cancer and Treatment Modalities. Curr Cancer Ther Rev 2017; 13: 112.10.2174/1573394713666170531081818CrossRefGoogle Scholar
Talebi, AS, Hejazi, P, Jadidi, M, Ghorbani, RA. Monte Carlo study on Photoneutron Spectrum around Elekta SL75/25 18 MV linear accelerator. Int J Adv Biol Biomed Res 2016; 4: 4857.Google Scholar
Rajesh, KR, Ganapathi Raman, R, Musthafa, MM, Midhun, CV, Joseph, N. A passive method for absolute dose evaluation of photoneutrons in radiotherapy. Int J Radiat Res 2020; 18: 173178.Google Scholar
Naseri, A, Mesbahi, A. A review on photoneutrons characteristics in radiation therapy with high-energy photon beams. Rep Pract Oncol Radiother 2010; 15: 138144.10.1016/j.rpor.2010.08.003CrossRefGoogle ScholarPubMed
Parwaie, W, Refahi, S, Afkhami Ardekani, M, Farhood, B. Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons. J Med Signals Sens 2018; 8: 195203.Google ScholarPubMed
Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA–AAPM international Code of Practice for reference and relative dose determination. Med Phys 2018; 45: 11231145.Google Scholar
Hosseinzadeh, E, Banaee, N, Nedaie, HA. Monte Carlo calculation of photo-neutron dose produced by circular cones at 18 MV photon beams. Rep Pract Oncol Radiother 2018; 23: 3946.10.1016/j.rpor.2017.12.001CrossRefGoogle ScholarPubMed
Samiei, F, Nedaie, HA, Darestani, H, Banaee, N, Shagholi, N, Mohammadi, K, Shahvar, A. The measurement of photoneutron dose around an 18 MV Varian linear accelerator by TLD600 and TLD700 dosimeters inside the polyethylene sphere of neutron probe LB 6411. Adv Phys 2014; 6: 10491057.Google Scholar
Saeed, MK, Moustafa, O, Yasin, OA, Tuniz, C, Habbani, FI. Doses to patients from photoneutrons emitted in a medical linear accelerator. Radiati Prot Dosimetry 2009; 133: 130135.10.1093/rpd/ncp029CrossRefGoogle Scholar
Zh, Guoqing. IAEA Germany, Monte Carlo simulation of mixed neutron-gamma radiation fields and dosimetry devices. Dec 2011; 150 p; Diss. (Dr.-Ing.) (2011).Google Scholar
Allahverdi, M, Zabihzadeh, M, Ay, MA, Mahdavi, SR, Shahriari, M, Mesbahi, A, Alijanzadeh, H. Monte Carlo estimation of electron contamination in a 18 MV clinical photon beam. Iran J Radiat Res 2011; 9: 1528.Google Scholar
Fielding, A. Monte-Carlo techniques for radiotherapy applications I: Introduction and overview of the different Monte-Carlo codes. J Radiother Pract 2023; 22: E80.10.1017/S1460396923000079CrossRefGoogle Scholar
Haghighat, A, Wanger, JC. Monte Carlo variance reduction with deterministic importance function. Prog Nucl Energy 2003; 42: 2553.10.1016/S0149-1970(02)00002-1CrossRefGoogle Scholar
Martínez-Ovalle, SA, Barquero, R, Gómez-Ros, JM, Lallena, AM. Neutron dose equivalent and neutron spectra in tissue for clinical linacs operating at 15, 18 and 20 MV. Radiat Prot Dosimetry 2011; 147: 498511.10.1093/rpd/ncq501CrossRefGoogle ScholarPubMed
Ghorbani, M, Vejdani Noghreiyan, A, Tabatabaei, ZS, Pakravan, D, Davenport, D. Tissue composition effect on dose distribution in radiotherapy with a 6 MV photon beam of a medical linac. J Cancer Res Ther 2019; 15: 237244.Google ScholarPubMed
Bahreyni Toossi, MT, Khajetash, B, Ghorbani, M. Assessment of Neutron Contamination Originating from the Presence of Wedge and Block in Photon Beam Radiotherapy. J Biomed Phys Eng 2018; 8: 312.Google ScholarPubMed
Kry, S F, Howell, R M, Salehpour, M, Followill, DS. Neutron spectra and dose equivalents calculated in tissue for high-energy radiation therapy. Med Phys 2009; 36: 1244–50.10.1118/1.3089810CrossRefGoogle ScholarPubMed
Zanini, A, Durisi, E, Fasolo, F, Ongaro, C, Visca, L, Nastasi, U, Burn, KW, Scielzo, G, Adler, JO, Annand, JR, Rosner, G. Monte Carlo simulation of the photoneutron field in LINAC radiotherapy treatments with different collimation systems. Phys Med Biol 2004; 49: 571582.10.1088/0031-9155/49/4/008CrossRefGoogle ScholarPubMed
Tajiki, S, Nedaie, HA, Rahmani, F. A Monte Carlo study of neutron contamination in presence of circular cones during stereotactic radiotherapy with 18MV photon beams. Biomed Phys Eng Express 2020; 6: 133.10.1088/2057-1976/ab7ff2CrossRefGoogle Scholar
Rezaian, A, Nedaie, HA, Banaee, N. Measurement of neutron dose in the compensator IMRT treatment. Appl Radiat Isot 2017; 128: 136141.10.1016/j.apradiso.2017.06.013CrossRefGoogle ScholarPubMed