Hostname: page-component-7479d7b7d-pfhbr Total loading time: 0 Render date: 2024-07-12T07:54:49.434Z Has data issue: false hasContentIssue false

Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps

Published online by Cambridge University Press:  01 July 2007

S. B. Neff
Affiliation:
University Children’s Hospital Zurich, Department of Anaesthesia, Zurich, Switzerland University Hospital of Zurich, Department of Anaesthesia, Zurich, Switzerland
T. A. Neff
Affiliation:
University Hospital of Zurich, Department of Anaesthesia, Zurich, Switzerland
S. Gerber
Affiliation:
University Children’s Hospital Zurich, Department of Anaesthesia, Zurich, Switzerland
M. M. Weiss*
Affiliation:
University Children’s Hospital Zurich, Department of Anaesthesia, Zurich, Switzerland
*
Correspondence to: Markus Weiss, Department of Anaesthesia, University Children’s Hospital, Steinwiesstrasse 75, CH-8032 Zurich, Switzerland. E-mail: markus.weiss@kispi.unizh.ch; Tel: +41 44 266 7111; Fax: +41 44 266 7449
Get access

Summary

Background and objective

Significant start-up delays are inherent to syringe infusion pumps, particularly at low infusion rates, as routinely used in children’s anaesthesia and intensive care. Such delays are mainly the result of engagement of gears in the mechanical drive or compliance of the syringe assembly. The purpose of the present study was to determine the effect of flow rate, syringe size and syringe architecture on fluid delivery during infusion start-up.

Methods

Elapsed time from infusion start to achievement of steady-state flow was gravimetrically determined for various infusion rates (0.1, 0.5, 1 mL h−1), different syringe sizes (10-, 20-, 30-, 50-mL) and syringes of two different brands (BD and Codan). Four measurements for each condition were performed with two identical Alaris® Asena™ GH syringe infusion pumps (total of eight experiments). Statistical analysis was done by two-way ANOVA with Bonferroni’s post-test; P < 0.05 was considered significant.

Results

Start-up time was from 3.6 ± 0.9 min (BD 10-mL syringe, 1.0 mL h−1) to 74.5 ± 26.6 min (BD 50-mL syringe, 0.1 mL h−1). Overall, the start-up time markedly increased with lower flow rate (0.1 mL h−1 vs. 1 mL h−1; P < 0.0001), larger syringe size (50 mL vs. 10 mL; P < 0.01), and the BD brand in comparison with the Codan syringes (P < 0.01).

Conclusions

Highest possible flow rate, smaller sized syringes and syringe plungers with reduced compressibility should be preferred in order to avoid significant start-up delays in fluid delivery.

Type
Original Article
Copyright
Copyright © European Society of Anaesthesiology 2007

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1.Rooke, GA, Bowdle, TA. Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures. Anesth Analg 1994; 78: 150156.CrossRefGoogle ScholarPubMed
2.Shibata, H, Aibiki, M, Shirakawa, Y, Ogli, K. Dopamine infused continuously at high concentration with a low flow rate affects arterial blood pressure fluctuation waves. Crit Care Med 1993; 21: 801804.CrossRefGoogle ScholarPubMed
3.Klem, SA, Farrington, JM, Leff, RD. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit Care Med 1993; 21: 12131217.CrossRefGoogle ScholarPubMed
4.Krauskopf, KH, Rauscher, J, Brandt, L. Disturbance of continuous, pump administration of cardiovascular drugs by hydrostatic pressure. Anaesthesist 1996; 45: 449452.CrossRefGoogle ScholarPubMed
5.Schulze, KF, Graff, M, Schimmel, MS, Schenkman, A, Rohan, P. Physiologic oscillations produced by an infusion pump. J Pediatr 1983; 103: 796798.Google Scholar
6.Leff, RD, Roberts, RJ. Problems in drug therapy for pediatric patients. Am J Hosp Pharm 1987; 44: 865870.Google Scholar
7.Capes, DF, Dunster, KR, Sunderland, VB, McMillan, D, Colditz, PB, McDonald, C. Fluctuations in syringe-pump infusions: association with blood pressure variations in infants. Am J Health Syst Pharm 1995; 52: 16461653.CrossRefGoogle ScholarPubMed
8.Lonnqvist, PA, Lofqvist, B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intens Care Med 1997; 23: 9981001.Google ScholarPubMed
9.Cunningham, S, Deere, S, McIntosh, N. Cyclical variation of blood pressure and heart rate in neonates. Arch Dis Child 1993; 69: 6467.CrossRefGoogle ScholarPubMed
10.Lonnqvist, PA. How continuous are continuous drug infusions? Intens Care Med 2000; 26: 660661.Google Scholar
11.Weiss, M, Hug, MI, Neff, T, Fischer, J. Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anaesth 2000; 47: 10311035.CrossRefGoogle ScholarPubMed
12.Weiss, M, Fischer, J, Neff, T, Baenziger, O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 2000; 55: 10941098.CrossRefGoogle ScholarPubMed
13.Weiss, M, Banziger, O, Neff, T, Fanconi, S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intens Care Med 2000; 26: 776779.CrossRefGoogle ScholarPubMed
14.Neff, TA, Fischer, JE, Schulz, G, Baenziger, O, Weiss, M. Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intens Care Med 2001; 27: 287291.CrossRefGoogle ScholarPubMed
15.Kern, H, Kuring, A, Redlich, U et al. . Downward movement of syringe pumps reduces syringe output. Br J Anaesth 2001; 86: 828831.Google Scholar
16.Igarashi, H, Obata, Y, Nakajima, Y, Katoh, T, Morita, K, Sato, S. Syringe pump displacement alters line internal pressure and flow. Can J Anaesth 2005; 52: 685691.Google Scholar
17.Cook, RI. Syringe pump assemblies and the natural history of clinical technology. Can J Anaesth 2000; 47: 929935.Google Scholar
18.Neff, T, Fischer, J, Fehr, S, Baenziger, O, Weiss, M. Start-up delays of infusion syringe pumps. Paediatr Anaesth 2001; 11: 561565.CrossRefGoogle ScholarPubMed
19.Leff, RD, True, WR, Roberts, RJ. A gravimetric technique for evaluating flow continuity from two infusion devices. Am J Hosp Pharm 1987; 44: 13881391.Google Scholar
20.Stull, JC, Erenberg, A, Leff, RD. Flow rate variability from electronic infusion devices. Crit Care Med 1988; 16: 888891.Google Scholar
21.Schulz, G, Fischer, J, Neff, T, Banziger, O, Weiss, M. The effect of air within the infusion syringe on drug delivery of syringe pump infusion systems. Anaesthesist 2000; 49: 10181023.CrossRefGoogle ScholarPubMed
22.McCarroll, C, McAtamney, D, Taylor, R. Alteration in flow delivery with antisyphon devices. Anaesthesia 2000; 55: 355357.Google Scholar
23.Neff, T, Fischer, J, Fehr, S, Baenziger, O, Weiss, M. Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems. Swiss Med Wkly 2001; 131: 219222.Google ScholarPubMed
24.Amoore, J, Dewar, D, Ingram, P, Lowe, D. Syringe pumps and start-up time: ensuring safe practice. Nurs Stand 2001; 15: 4345.CrossRefGoogle ScholarPubMed
25.Powell, ML, Carnevale, FA. A comparison between single and double-pump syringe changes of intravenous inotropic medications in children. Dynamics 2004; 15: 1014.Google Scholar
26.Trim, JC, Roe, J. Practical considerations in the administration of intravenous vasoactive drugs in the critical care setting: the double pumping or piggyback technique-part one. Intens Crit Care Nurs 2004; 20: 153160.CrossRefGoogle ScholarPubMed
27.Kim, DW, Steward, DJ. The effect of syringe size on the performance of an infusion pump. Paediatr Anaesth 1999; 9: 335337.CrossRefGoogle ScholarPubMed
28.Weiss, M, Neff, T, Gerber, A, Fischer, J. Impact of infusion line compliance on syringe pump performance. Paediatr Anaesth 2000; 10: 595599.Google Scholar
29.Donmez, A, Araz, C, Kayhan, Z. Syringe pumps take too long to give occlusion alarm. Paediatr Anaesth 2005; 15: 293296.Google Scholar
30.Weiss, M, Gerber, S, Fuchslin, RM, Neff, TA. Accurate continuous drug delivery at low infusion rate with a novel microvolumetric infusion pump (MVIP): pump design, evaluation and comparison to the current standard. Anaesthesia 2004; 59: 11331137.Google Scholar