Hostname: page-component-77c89778f8-m42fx Total loading time: 0 Render date: 2024-07-21T00:36:22.602Z Has data issue: false hasContentIssue false

Sensitivity and Detectability Limits for Elemental Analysis by Proton-Induced X-Ray Fluorescence with a 3 MV Van De Graaff*

Published online by Cambridge University Press:  06 March 2019

C. J. Umbarger
Affiliation:
University of California, Los Alamos Scientific Laboratory, Los Alamos, New Mexico 87544
R. C. Bearse
Affiliation:
University of California, Los Alamos Scientific Laboratory, Los Alamos, New Mexico 87544
D. A. Close
Affiliation:
University of California, Los Alamos Scientific Laboratory, Los Alamos, New Mexico 87544
J. J. Malanify
Affiliation:
University of California, Los Alamos Scientific Laboratory, Los Alamos, New Mexico 87544
Get access

Abstract

Protons from a 3 MV Van de Graaff have been used to produce characteristic x-rays from 21 elements spanning the periodic table. Absolute Kα and Lα x-ray production cross sections have been determined, allowing one to calculate sensitivities for any given sample, detector geometry, and proton beam parameters. Elemental detectability limits are discussed assuming various backings (e.g., mylar, kapton, etc.) and matrix materials. The large number of available small proton accelerators throughout the United States promises wide applicability of this technique to environmental and biomedical analysis.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1972

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

Footnotes

Visiting Staff Member from University of Kansas, Lawrence, Kansas.

*

A Work performed under the auspices of the U. S. Atomic Energy Commission.

References

1. Flocchini, R. G., Feeney, P. J., Sommerville, R. J., and Cahill, T. A., “Sensitivity Versus Target Backings for Elemental Analysis by Alpha Excited X-Ray Emission,Nucl. Instr. Meth. 100, 397402 (1972).Google Scholar
2. Watson, R. L., Sjurseth, J. R., and Howard, R. W., “An Investigation of the Analytical Capabilities of X-Ray Emission Induced by High Energy Alpha Particles,” Nucl. Instr. Meth. 93, 6976 (1971).Google Scholar
3. Rudolph, H., Ristinen, R. A., Smyth, W. R., Alfrey, A., and Contiguglia, S. R., “A Proton-Induced X-Ray Study of Trace Element Transfer in Hemodialysis,Bull. Am. Phys. Soc. 17, 548 (1972).Google Scholar
4. Johansson, T. B., Akselsson, R., and Johansson, S. A. E., “X-Ray Analysis: Elemental Trace Analysis at the 10-12g Level,Nucl. Instr. Meth. 84, 141143 (1970).Google Scholar
5. Bissinger, G. A., Shafroth, S. M., and Waltner, A. W., “Yields of K and L X-Rays Arising From 2-30 MeV Proton Bombardment of Ag,Phys. Rev. A 5, 20462052 (1972).Google Scholar
6. Bissinger, G. A., Hoyce, J. M., Ludwig, E. J., McEver, W. S., and Shafroth, S. M., “Study of the Production of K X-Rays in Ca, Ti, and Ni by 2-28 MeV Protons,” Phys. Rev. A X, 841847 (1970).Google Scholar
7. Merzbacher, E and Lewis, H. W., “X-Ray Production by Heavy Charged Particles,” in Flügge, S., Editor, Encyclopedia of Physics, Vol. 34, 166192, Springer, Berlin (1958).Google Scholar
8. Lewis, H. W., Simmons, B. E., and Merzbacher, E., “Production of Characteristic X-Rays by Protons of 1.7 to 3.0 MeV Energy,Phys. Rev. 91, 943946 (1953).Google Scholar
9. Khan, J. M., Potter, D. L., and Worley, R. D., “Studies in X-Ray Production by Proton Bombardment of C, Mg, Al, Nd, Sm, Gd, Tb, Dy, and Ho,” Phys. Rev. 139, A 1735–46 (1965).Google Scholar
10. Micro Matter Co., 197 34th St. East, Seattle, Washington 98102.Google Scholar
11. Feldl, E. J. and Umbarger, C. J., “A Target Chamber for Charged Particle Induced X-Ray Fluorescence Studies,” Nucl. Instr. Meth. (to be published).Google Scholar
12. Golovnya, V. Y., Klyucharev, A. P., Shilyaev, E. A., and Shlyakhov, N. A., “Elastic Scattering of Low-Energy Protons by Isotopes of Cr, Fe, Ni, and Co,” Sov. J. of Nucl. Phys. 1 3236 (1965).Google Scholar
13. Ortec 7000 Series Detector Manual, Ortec Inc., 100 Midland Rd., Oak Ridge, Tennessee 37830.Google Scholar