Feasibility and Accuracy of a Wearable Biosensor Device for Vital Sign Monitoring in Septic Emergency Department Patients in Rwanda

  • Stephanie Garbern (a1), Gabin Mbanjumucyo (a2), Christian Umuhoza (a2), Vinay Sharma (a3), James Mackey (a4), Kyle Martin (a1), Francois Twagirumukiza (a2), Samantha Rosman (a5), Natalie McCall (a6), Stephan Wegerich (a7) and Adam Levine (a1)...

Abstract

Introduction:

Low and middle-income countries (LMICs) bear a disproportionately high burden of sepsis, contributing to an estimated 90% of global sepsis-related deaths. Critical care capabilities needed for septic patients, such as continuous vital sign monitoring, are often unavailable in LMICs.

Aim:

This study aimed to assess the feasibility and accuracy of using a small wireless, wearable biosensor device linked to a smartphone, and a cloud analytics platform for continuous vital sign monitoring in emergency department (ED) patients with suspected sepsis in Rwanda.

Methods:

This was a prospective observational study of adult and pediatric patients (≥ 2 months) with suspected sepsis presenting to Kigali University Teaching Hospital ED. Biosensor devices were applied to patients’ chest walls and continuously recorded vital signs (including heart rate and respiratory rate) for the duration of their ED course. These vital signs were compared to intermittent, manually-collected vital signs performed by a research nurse every 6-8 hours. Pearson’s correlation coefficients were calculated over the study population to determine the correlation between the vital signs obtained from the biosensor device and those collected manually.

Results:

42 patients (20 adults, 22 children) were enrolled. Mean duration of monitoring with the biosensor device was 34.4 hours. Biosensor and manual vital signs were strongly correlated for heart rate (r=0.87, p<0.001) and respiratory rate (r=0.74 p<0.001). Feasibility issues occurred in 9/42 (21%) patients, although were minor and included biosensor falling off (4.8%), technical/connectivity problems (7.1%), removal by a physician (2.4%), removal for a procedure (2.4%), and patient/parent desire to remove the device (4.8%).

Discussion:

Wearable biosensor devices can be feasibly implemented and provide accurate continuous vital sign measurements in critically ill pediatric and adult patients with suspected sepsis in a resource-limited setting. Further prospective studies evaluating the impact of biosensor devices on improving clinical outcomes for septic patients are needed.

    • Send article to Kindle

      To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

      Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

      Find out more about the Kindle Personal Document Service.

      Feasibility and Accuracy of a Wearable Biosensor Device for Vital Sign Monitoring in Septic Emergency Department Patients in Rwanda
      Available formats
      ×

      Send article to Dropbox

      To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

      Feasibility and Accuracy of a Wearable Biosensor Device for Vital Sign Monitoring in Septic Emergency Department Patients in Rwanda
      Available formats
      ×

      Send article to Google Drive

      To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

      Feasibility and Accuracy of a Wearable Biosensor Device for Vital Sign Monitoring in Septic Emergency Department Patients in Rwanda
      Available formats
      ×

Copyright

Recommend this journal

Email your librarian or administrator to recommend adding this journal to your organisation's collection.

Prehospital and Disaster Medicine
  • ISSN: 1049-023X
  • EISSN: 1945-1938
  • URL: /core/journals/prehospital-and-disaster-medicine
Please enter your name
Please enter a valid email address
Who would you like to send this to? *
×

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed