Intra-operative skull X-ray for misdirection of the cochlear implant array into the vestibular labyrinth

  • A M Hassan (a1), R Patel (a2) and M Redleaf (a3)

Abstract

Objectives:

This paper reports five cases of aberrant cochlear implant electrode array insertion into the vestibular labyrinth. A review of the literature was conducted in order to clarify reasonable preventive and detection strategies and endorse the routine use of intra-operative plain skull X-ray.

Methods:

The study entailed a clinical case series and literature review. The setting was a tertiary academic referral centre. The following data were evaluated: pre-operative temporal bone computed tomography, operative reports, intra-operative imaging, neural response telemetry/imaging and post-operative imaging.

Results:

There were no consistent pre-operative risk factors found on computed tomography scans and no reliable intra-operative signs of electrode array misdirection. All misdirections in our case series, and those in the literature, were easily detectable on intra-operative plain film X-ray.

Conclusion:

These reported cases demonstrate implant misdirection without the surgeon's awareness. Aberrant insertion cannot be anticipated, and neural response telemetry/imaging is not a reliable indicator of misdirection. Routine intra-operative anteroposterior plain X-ray of the head is a reliable indicator of misdirection, and is fast and relatively inexpensive.

Copyright

Corresponding author

Address for correspondence: Mr Ahmad M Hassan, College of Medicine, University of Illinois at Chicago, 808 S Wood St, Chicago, IL 60612, USA E-mail: ahassa8@uic.edu

References

Hide All
1National Institute on Deafness and Other Communication Disorders (NIDCD). Cochlear Implants. In: http://www.nidcd.nih.gov/health/hearing/coch.asp [3 December 2012]
2Carlson, ML, Driscoll, CL, Gifford, RH, McMenomey, SO. Cochlear implantation: current and future device options. Otolaryngol Clin North Am 2012;45:221–48
3Magnus, B, Rivas, A, Tsai, BS, Haynes, DS, Roland, JT. Surgical techniques in cochlear implants. Otolaryngol Clin North Am 2012;45:6980
4Flint, PW, Haughey, BH, Niparko, JK, Richardson, MA, Lund, VJ, Robbins, KT et al. , eds. Cummings Otolaryngology – Head and Neck Surgery, 5th edn.Philadelphia: Mosby/Elsevier, 2010;2237
5Cohen, NL, Roland, JT, Fishman, A. Surgical technique for the nucleus contour cochlear implant. Ear Hear 2002;23(1 suppl):5966S
6Balkany, TJ, Whitley, M, Shapira, Y, Angeli, S, Brown, K, Eter, E et al. The temporalis pocket technique for cochlear implantation: an anatomic and clinical study. Otol Neurotol 2009;30:903–7
7MED-EL Surgical Manual, version 1. Durham, NC: US Med-El, 1998;15
8Copeland, BJ, Pillsbury, HC, Buchman, CA. Prospective evaluation of intraoperative cochlear implant radiographs. Otol Neurotol 2004;25:295–7
9Coelho, DH, Waltzman, SB, Roland, JT Jr. Implanting common cavity malformations using intraoperative fluoroscopy. Otol Neurotol 2008;29:914–19
10Fishman, AJ, Roland, JT Jr, Alexiades, G, Mierzwinski, J, Cohen, NL. Fluoroscopically assisted cochlear implantation. Otol Neurotol 2003;24:882–6
11Yuan, YY, Song, YS, Chai, CM, Shen, WD, Han, WJ, Liu, J et al. Intraoperative CT-guided cochlear implantation in congenital ear deformity. Acta Otolaryngol 2012;132:951–8
12Shpizner, BA, Holliday, RA, Roland, JT, Cohen, NL, Waltzman, SB, Shapiro, WH. Postoperative imaging of the multichannel cochlear implant. AJNR Am J Neuroradiol 1995;16:1517–24
13Shapiro, WH, Bradham, TS. Cochlear implant programming. Otolaryngol Clin North Am 2012;45:111–27
14Mason, S. Electrophysiologic and objective monitoring of the cochlear implant during surgery: implementation, audit, and outcomes. Int J Audiol 2004;43:S338
15Tange, RA, Grolman, W, Maat, A. Intracochlear misdirected implantation of a cochlear implant. Acta Otolaryngol 2006;126:650–2
16Fishman, A. Imaging and anatomy for cochlear implants. Otolaryngol Clin North Am 2012;45:124
17Mecca, MA, Wagle, W, Lupinetti, A, Parnew, S. Complication of cochlear implant surgery. AJNR Am J Neuroradiol 2003;24:2089–91
18Muzzi, E, Boscolo-Rizzo, P, Santarelli, R. Cochlear implant electrode array misplacement: a cautionary case report. J Laryngol Otol 2012;126:414–17
19Ramalingam, R, Ramalingam, KK, Padmaja, HS. An unusual occurrence in cochlear implant surgery: misplaced electrode. J Laryngol Otol 2009;123:e4
20Rotteveel, LJ, Proops, DW, Ramsden, RT. Cochlear implantation in 53 patients with otosclerosis: demographics, computed tomographic scanning, surgery, and complications. Otol Neurotol 2004;25:943–52
21Viccaro, M, Covelli, E, De Seta, E, Balsamo, G, Filipo, R. The importance of intra-operative imaging during cochlear implant surgery. Cochlear Implants Int 2009;10:198202
22Sorrentino, T, Cote, M, Eter, E, Laborde, ML, Cochard, N, Deguine, O et al. Cochlear reimplantations: technical and surgical failures. Acta Otolaryngol 2009;129:380–4
23Ying, YM, Lin, JW, Oghalai, JS, Williamson, RA. Cochlear implant electrode misplacement: incidence, evaluation and management. Laryngoscope 2013;123:757–66
24Cosetti, MK, Troob, SH, Latzman, JM, Shapiro, WH, Roland, JT Jr, Waltzman, SB. An evidence-based algorithm for intraoperative monitoring during cochlear implantation. Otol Neurotol 2012;33:169–76

Keywords

Metrics

Altmetric attention score

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed