3 - Spectral analysis methods, quantitation, and common artifacts

Summary

Key points

Correct post-processing and quantitation are key aspects of in vivo MRS.

Filtering, phase-correction and baseline correction improve MRS data.

Peak area estimation can be done using parametric or non-parametric routines in either the time domain or frequency domains.

“LCModel” software is becoming widely used and accepted, particularly for single-voxel MRS data.

MRSI processing requires additional steps; k-space filtering and other manipulations can improve MRSI data quality.

A variety of strategies are available for quantitation, based on either internal or external reference standards, or phantom replacement methodology.

Quantitation routines should take into account voxel composition, particularly the amount of CSF partial volume present.

MRS is sensitive to field inhomogeneity and other artifacts.

Introduction

Methods for spectral analysis and the quantitative analysis of spectral data are arguably as important as the techniques used to collect the data; the use of incorrect analysis methods can lead to systematic errors or misinterpretation of spectra. In general, the ultimate goal of spectral analysis is to determine the concentrations of the compounds present in the spectra. In MRS, the area under the spectral peak is proportional to the metabolite concentration; however, determining the proportionality constant can be challenging. In addition, peak area measurements in in-vivo spectroscopy are complicated by resonance overlap, baseline distortions, and lineshapes that often only poorly approximate conventional models such as Gaussian or Lorentzian functions. Therefore, quantitative analysis of in vivo MRS data is challenging. This chapter reviews basic spectral processing techniques, methods for determining peak areas, and strategies for calculating metabolite concentrations.

References
Ordidge, RJ, Cresshull, ID. The correction of transient B0 field shifts following the application of pulsed gradients by phase correction in the time domain. J Magn Reson Imaging 1986; 69: 151–5.
Graaf, AA, Dijk, JE, Bovee, WM. QUALITY: quantification improvement by converting lineshapes to the Lorentzian type. Magn Reson Med 1990; 13: 343–57.
Marion, D, Ikura, M, Bax, A. Improved solvent suppression in one- and two-dimensional NMR spectra by convolution of time-domain data. J Magn Reson 1989; 84: 425–30.
Lowe, MJ, Sorenson, JA. Spatially filtering functional magnetic resonance imaging data. Magn Reson Med 1997; 37: 723–9.
Derby, K, Hawryszko, C, Tropp, J. Baseline deconvolution, phase correction, and signal quantification in Fourier localized spectroscopic imaging. Magn Reson Med 1989; 12: 235–40.
Soher, BJ, Young, K, Govindaraju, V, Maudsley, AA. Automated spectral analysis III: application to in vivo proton MR spectroscopy and spectroscopic imaging. Magn Reson Med 1998; 40: 822–31.
Soher, BJ, Zijl, PC, Duyn, JH, Barker, PB. Quantitative proton MR spectroscopic imaging of the human brain. Magn Reson Med 1996; 35:356–63.
Mierisova, S, Ala-Korpela, M. MR spectroscopy quantitation: a review of frequency domain methods. NMR Biomed 2001; 14: 247–59.
Marshall, I, Bruce, SD, Higinbotham, J, MacLullich, A, Wardlaw, JM, Ferguson, KJ, et al. Choice of spectroscopic lineshape model affects metabolite peak areas and area ratios. Magn Reson Med 2000; 44: 646–9.
Soher, BJ, Maudsley, AA. Evaluation of variable line-shape models and prior information in automated 1H spectroscopic imaging analysis. Magn Reson Med 2004; 52: 1246–54.
Provencher, SW. Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med 1993;30: 672–9.
Beer, R, Boogaart, AOrmondt, D, Pijnappel, WW, Hollander, JA, Marien, AJ, et al. Application of time-domain fitting in the quantification of in vivo 1H spectroscopic imaging data sets. NMR Biomed 1992; 5: 171–8.
Barker, PB, Soher, BJ, Blackband, SJ, Chatham, JC, Mathews, VP, Bryan, RN. Quantitation of proton NMR spectra of the human brain using tissue water as an internal concentration reference. NMR Biomed 1993; 6: 89–94.
Freeman, R. A Handbook of Nuclear Magnetic Resonance. Harlow, England: Longman Scientific and Technical, 1987.
Hetherington, HP, Mason, GF, Pan, JW, Ponder, SL, Vaughan, JT, Twieg, DB, et al. Evaluation of cerebral gray and white matter metabolite differences by spectroscopic imaging at 4.1 T. Magn Reson Med 1994; 32: 565–71.
Stockler, S, Holzbach, U, Hanefeld, F, Marquardt, I, Helms, G, Requart, M, et al. Creatine deficiency in the brain: a new, treatable inborn error of metabolism. Pediatr Res 1994; 36: 409–13.
Thulborn, KR, Ackerman, JJH. Absolute molar concentrations by NMR in inhomogeneous B1. A scheme for analysis of in vivo metabolites. J Magn Reson 1983; 55: 357–71.
Christiansen, P, Henriksen, O, Stubgaard, M, Gideon, P, Larsson, HB. In vivo quantification of brain metabolites by 1H-MRS using water as an internal standard. Magn Reson Imaging 1993; 11: 107–18.
Alger, JR, Symko, SC, Bizzi, A, Posse, S, DesPres, DJ, Armstrong, MR. Absolute quantitation of short TE brain 1H-MR spectra and spectroscopic imaging data. J Comput Ass Topogr 1993; 17:191–9.
Ordidge, RJ, Cresshull, ID. The correction of transient B0 field shifts following the application of pulsed gradients by phase correction in the time domain. J Magn Reson 1986; 69: 151–5.
Webb, PG, Sailasuta, N, Kohler, SJ, Raidy, T, Moats, RA, Hurd, RE. Automated single-voxel proton MRS: technical development and multisite verification. Magn Reson Med 1994; 31: 365–73.
Soher, BJ, Hurd, RE, Sailasuta, N, Barker, PB. Quantitation of automated single-voxel proton MRS using cerebral water as an internal reference. Magn Reson Med 1996; 36: 335–9.
Mathews, VP, Barker, PB, Blackband, SJ, Chatham, JC, Bryan, RN. Cerebral metabolites in patients with acute and subacute strokes: concentrations determined by quantitative proton MR spectroscopy. Am J Radiol 1995; 165: 633–8.
Ernst, T, Kreis, R, Ross, B. Absolute quantitation of water and metabolites in the human brain. I. Compartments and water. J Magn Reson B 1993; 102: 1–8.
Horská, A, Calhoun, VD, Bradshaw, DH, Barker, PB. Rapid method for correction of CSF partial volume in quantitative proton MR spectroscopic imaging. Magn Reson Med 2002; 48: 555–8.
Horska, A, Jacobs, MA, Calhoun, V, Arslanoglu, A, Barker, PB. A fast method for image segmentation: application to quantitative proton MRSI at 3 Tesla. ISMRM, 11th Scientific Meeting and Exhibition; Toronto, Canada; 2003.
Ernst, T, Chang, L. Elimination of artifacts in short echo time H MR spectroscopy of the frontal lobe. Magn Reson Med 1996; 36: 462–8.
Hurd, R, Sailasuta, N. Elimination of artifacts in short echo proton spectroscopy. 5th Annual Meeting of the International Society of Magnetic Resonance in Medicine;Vancouver, BC, Canada; 1997. p. 1453.
Murdoch, JM. Still Iterating … and iterating … to solve pulse design problems. International Society of Magnetic Resonance in Medicine (ISMRM) 10th Annual Meeting; Honolulu, Hawai'i; 2002. p. 923.
Edden, RA, Barker, PB. Spatial effects in the detection of gamma-aminobutyric acid: improved sensitivity at high fields using inner volume saturation. Magn Reson Med 2007; 58: 1276–82.
Edden, RA, Schar, M, Hillis, AE, Barker, PB. Optimized detection of lactate at high fields using inner volume saturation. Magn Reson Med 2006; 56: 912–7.
Tran, TK, Vigneron, DB, Sailasuta, N, Tropp, J, Roux, P, Kurhanewicz, J, et al. Very selective suppression pulses for clinical MRSI studies of brain and prostate cancer. Magn Reson Med 2000; 43: 23–33.
Lin, AP, Ross, BD. Short-echo time proton MR spectroscopy in the presence of gadolinium. J Comput Assist Tomogr 2001; 25: 705–12.
Sijens, PE, Oudkerk, M, Dijk, P, Levendag, PC, Vecht, CJ. 1H MR spectroscopy monitoring of changes in choline peak area and line shape after Gd-contrast administration. Magn Reson Imaging 1998; 16: 1273–80.
Sijens, PE, Bent, MJ, Nowak, PJ, Dijk, P, Oudkerk, M. 1H chemical shift imaging reveals loss of brain tumor choline signal after administration of Gd-contrast. Magn Reson Med 1997; 37: 222–5.
Windham, JP, Abd-Allah, MA, Reimann, DA, Froelich, JW, Haggar, AM. Eigenimage filtering in MR imaging. J Comput Assist Tomogr 1988; 12: 1–9.
Kreis, R. Issues of spectral quality in clinical 1H-magnetic resonance spectroscopy and a gallery of artifacts. NMR Biomed 2004; 17: 361–81.