Hostname: page-component-8448b6f56d-c47g7 Total loading time: 0 Render date: 2024-04-18T06:55:46.277Z Has data issue: false hasContentIssue false

FloTrac for monitoring arterial pressure and cardiac output during phaeochromocytoma surgery

Published online by Cambridge University Press:  01 September 2008

O. Collange*
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
L. Xavier
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
H. Kuntzman
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
B. Calon
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
R. Schaeffer
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
T. Pottecher
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
P. Diemunsch
Affiliation:
Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
P. Pessaux
Affiliation:
Service de chirurgie générale, Hépatique et endocrinienne, Hôpitaux Universitaires de Strasbourg, Strasbourg, France
*
Correspondence to: Olivier Collange, Pôle d’Anesthésie, Réanimation Chirurgicale, SAMU, Hôpitaux Universitaires de Strasbourg, Hôpital de Hautepierre, Avenue Molière, 67098 Strasbourg, France. E-mail: olivier.collange@chrustrasbourg.fr; Tel: +33 388127075; Fax: +33 388127074

Abstract

Type
Correspondence
Copyright
Copyright © European Society of Anaesthesiology 2008

EDITOR:

The recently introduced FloTrac device (FloTrac/Vigileo; Edwards Lifescience, Irvine, CA, USA) has been proposed as a reliable semi-invasive device for cardiac output (CO) measurements. Its technology is based on arterial pressure waveform analysis and needs no calibration. We report its use during a haemodynamically unstable situation where the FloTrac device was compared to a pulmonary arterial catheter (PAC) for the measurement of CO.

Case report

A 34-yr-old female presented for surgical management of multiple endocrine neoplasia type IIa (MEN IIa). She was scheduled for elective bilateral adrenal tumour removal and total thyroidectomy. After induction of general anaesthesia, radial artery catheter and PAC (continuous CO 7.5 F catheter; Edwards Lifescience) were inserted. The FloTrac device was then attached to the arterial canula and connected to the Vigileo monitor (version 1.10). Concurrent recording of CO indexed to body surface area (cardiac index, CI) were performed on the FloTrac (CI_FloTrac) and the PAC (CI_PAC) during and after the surgery. Additional information collected included heart rate, arterial pressure, anaesthetic and surgical events. During phaeochromocytoma surgery, we observed an episode of severe hypertension (MAP = 140 mmHg) treated by deepening of anaesthesia and a bolus of urapidil (25 mg). After adrenalectomy, hypotension was treated with ephedrine boluses, neosynephrine and then norepinephrine (up to 0.6 μg kg−1 min−1) (Fig. 1).

Figure 1 CI_PAC, CI_FloTrac, MAP during and after phaeochromocytoma surgery. CI_PAC: cardiac index (CI) measured with the pulmonary arterial catheter in L min−1 m−2; CI_FloTrac: CI measured with the FloTrac device in L min−1 m−2; MAP: mean arterial pressure in mmHg. Arrows indicate durations of the different periods: Surgery, Phaeochromocytoma resection (‘Phaeo resec’), Recovery period.

Postoperative recovery was marked by the rapid resolution of arterial hypotension.

A total of 38 pairs of CI measurements were analysed. CI_FloTrac values ranged from 2.1 to 6.1 L min−1 m−2 (mean ± SD 3.4 ± 0.7 L min−1 m−2) and CI_PAC from 3.3 to 5.2 (mean 4 ± 0.4 L min−1 m−2). Simple regression and Bland–Altman analysis were used to compare the two methods of CI measurement. Simple regression was used to compare mean arterial pressure (MAP) and CI_FloTrac. The CI measured by the two methods did not correlate well (simple regression r 2 = 0.026, P = 0.33; Bland–Altman test r 2 = 0.23, bias = −0.66 L min−1 m−2, precision ± 0.75 L min−1 m−2, limits of agreement: −2.13 and +0.81 L min−1 m−2). CI_FloTrac was moderately correlated with the MAP: r 2 = 0.56, P < 0.001. Concurrent analysis of CI_FT, CI_PAC and MAP during phaeochromocytoma manipulation showed a dramatic increase in MAP (from 70 to 147 mmHg) and in CI_FloTrac (from 2.2 to 6.1 L min−1 m−2), whereas CI_PAC only slightly increased (from 3.3 to 4 L min−1 m−2).

Discussion

Phaeochromocytoma surgery usually necessitates standard and continuous arterial pressure monitoring. Measurement of pulmonary arterial occlusion pressure (PAOP) and CI with a PAC may be helpful to detect cardiac insufficiency and cardiogenic shock, particularly during manipulation of the tumour. PAC is usually proposed for patients with dramatic clinical manifestation and/or with preoperative cardiac dysfunction. In our case, we used a PAC considering the size of the bilateral tumours and the type of tumour (MEN IIa) overproducing all catecholamines. The use of the PAC has been increasingly criticized because it is an invasive technique and because of its unclear risk–benefit ratio [Reference Connors, Speroff and Dawson1Reference Shah, Hasselblad and Stevenson3]. Several haemodynamic devices have been developed using calculation of CI based on peripheral arterial pressure waveform analysis [Reference Rodig, Prasser, Keyl, Liebold and Hobbhahn4,Reference Yamashita, Nishiyama, Yokoyama, Abe and Manabe5]. These devices need to be initially calibrated by either transpulmonary thermodilution or pulmonary artery thermodilution to correct for arterial compliance in the calculation. The recently introduced FloTrac/Vigileo device calculates continuous CI on peripheral arterial pressure waveform characteristics but does not require external calibration. Individual data including height, weight, age, gender and the real-time arterial pressure waveform analysis are used to estimate arterial compliance. However, we found only five clinical studies assessing FloTrac as a CO monitor, most of them included cardiac surgery patients and compared CO measured by FloTrac and by PAC or PiCCO [Reference Manecke and Auger6Reference Sakka, Kozieras, Thuemer and van Hout10]. Two studies showed satisfactory correlation [Reference Manecke and Auger6,Reference Button, Weibel, Reuthebuch, Genoni, Zollinger and Hofer9], one study showed moderate correlation [Reference Mayer, Boldt, Schollhorn, Rohm, Mengistu and Suttner7] and two studies showed limited correlation [Reference Opdam, Wan and Bellomo8,Reference Sakka, Kozieras, Thuemer and van Hout10]. There is no data exploring FloTrac reliability during unstable haemodynamic situations such as haemorrhagic or catecholamine crisis.

We report the case of one patient for which 38 time points measurement of CI_FloTrac were compared to 38 measurements of CI_PAC. Of course, our data have to be interpreted very carefully and are insufficient to accurately define the global relationship between CI_FloTrac and CI_PAC. However, anaesthesists and intensivists have to tailor patient treatment individually and in our case, the FloTrac would have misled the physician on the patient’s haemodynamic status during the unstable period. Moreover, in our case report, CI_FloTrac values were better correlated to MAP than to CI_PAC values. The CI_FloTrac curve seemed to follow the huge increase in MAP during severe hypertension, whereas CI_PAC was only slightly increased. The FloTrac device needs demonstrative data before being considered as a monitor of CO for haemodynamically unstable patients.

References

1.JrConnors, AF, Speroff, T, Dawson, NV et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 1996; 276: 889897.CrossRefGoogle ScholarPubMed
2.Sandham, JD, Hull, RD, Brant, RF et al. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003; 348: 514.CrossRefGoogle ScholarPubMed
3.Shah, MR, Hasselblad, V, Stevenson, LW et al. Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 2005; 294: 16641670.CrossRefGoogle ScholarPubMed
4.Rodig, G, Prasser, C, Keyl, C, Liebold, A, Hobbhahn, J. Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 1999; 82: 525530.CrossRefGoogle ScholarPubMed
5.Yamashita, K, Nishiyama, T, Yokoyama, T, Abe, H, Manabe, M. Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anaesth (J Can Anesth) 2005; 52: 530534.CrossRefGoogle Scholar
6.JrManecke, GR, Auger, WR. Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 2007; 21: 37.CrossRefGoogle Scholar
7.Mayer, J, Boldt, J, Schollhorn, T, Rohm, KD, Mengistu, AM, Suttner, S. Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth 2007; 98: 176182.CrossRefGoogle ScholarPubMed
8.Opdam, HI, Wan, L, Bellomo, R. A pilot assessment of the FloTrac(TM) cardiac output monitoring system. Intensive Care Med 2007; 33: 344349.CrossRefGoogle Scholar
9.Button, D, Weibel, L, Reuthebuch, O, Genoni, M, Zollinger, A, Hofer, CK. Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 2007; 99: 329336.CrossRefGoogle ScholarPubMed
10.Sakka, SG, Kozieras, J, Thuemer, O, van Hout, N. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 2007; 99: 337342.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1 CI_PAC, CI_FloTrac, MAP during and after phaeochromocytoma surgery. CI_PAC: cardiac index (CI) measured with the pulmonary arterial catheter in L min−1 m−2; CI_FloTrac: CI measured with the FloTrac device in L min−1 m−2; MAP: mean arterial pressure in mmHg. Arrows indicate durations of the different periods: Surgery, Phaeochromocytoma resection (‘Phaeo resec’), Recovery period.