Hostname: page-component-55b6f6c457-kv5sj Total loading time: 0.598 Render date: 2021-09-26T15:07:21.403Z Has data issue: true Feature Flags: { "shouldUseShareProductTool": true, "shouldUseHypothesis": true, "isUnsiloEnabled": true, "metricsAbstractViews": false, "figures": true, "newCiteModal": false, "newCitedByModal": true, "newEcommerce": true, "newUsageEvents": true }

Inductive link for power and data transfer to a medical implant

Published online by Cambridge University Press:  04 August 2017

Giuseppina Monti*
Affiliation:
Department of Engineering for Innovation, University of Salento, Lecce, Italy. Phone: +39 0832 297365
Maria Valeria De Paolis
Affiliation:
Department of Engineering for Innovation, University of Salento, Lecce, Italy. Phone: +39 0832 297365
Laura Corchia
Affiliation:
Department of Engineering for Innovation, University of Salento, Lecce, Italy. Phone: +39 0832 297365
Mauro Mongiardo
Affiliation:
Department of Engineering, University of Perugia, Perugia, Italy
Luciano Tarricone
Affiliation:
Department of Engineering for Innovation, University of Salento, Lecce, Italy. Phone: +39 0832 297365
*
Corresponding author: G. Monti Email: giuseppina.monti@unisalento.it

Abstract

This paper presents a resonant inductive link for power and data transfer to a pulse generator implanted in the chest. The proposed link consists of two planar resonators and has been optimized for operating in the MedRadio band centered at 403 MHz. The wireless power/data link occurs between an external resonator operating in direct contact with the skin and a receiving resonator integrated in the silicone header of a pulse generator implanted in the chest. Numerical and experimental results are presented and discussed. From measurements performed by using minced pork to simulate the presence of human tissues, an efficiency of about 51% is demonstrated. The feasibility of using the proposed link for recharging the battery of the medical device in compliance with safety regulations is also verified and discussed.

Type
Wirelessly Powering: The Future
Copyright
Copyright © Cambridge University Press 2017 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

[1] Costanzo, A. et al. : Electromagnetic energy harvesting and wireless power transmission: a unified approach. Proc. IEEE, 102 (2014), 16921711.CrossRefGoogle Scholar
[2] Monti, G.; Corchia, L.; De Benedetto, E.; Tarricone, L.: A wearable wireless energy link for thin-film batteries charging. Int. J. Antennas Propag., 2016 (2016), 19.CrossRefGoogle Scholar
[3] Imura, T.; Okabe, H.; Hori, Y.: Basic experimental study on helical antennas of wireless power transfer for electric vehicles by using magnetic resonant couplings, in IEEE Vehicle Power and Propulsion Conf., Dearborn, 2009.Google Scholar
[4] Kurs, A.; Karalis, A.; Moffatt, R.; Joannopoulos, J.D.; Fisher, P.; Soljacic, M.: Wireless power transfer via strongly coupled magnetic resonances. Science, 317 (2007), 8386.CrossRefGoogle ScholarPubMed
[5] Monti, G.; Corchia, L.; Tarricone, L.: ISM band rectenna using a ring loaded monopole. Prog. Electromagn. Res. C, 33 (2012), 115.CrossRefGoogle Scholar
[6] Monti, G.; Corchia, L.; De Benedetto, E.; Tarricone, L.: Compact resonator on leather for nonradiative inductive power transfer and far-field data links. Radio Sci., 51 (2016), 629637.CrossRefGoogle Scholar
[7] Low, Z.N.; Chinga, R.A.; Tseng, R.; Lin, J.: Design and test of a high-power high-efficiency loosely coupled planar wireless power transfer system. IEEE Trans. Ind. Electron., 56 (2009), 18011812.Google Scholar
[8] Kuo, N.-C.; Zhao, B.; Niknejad, A.M.: Bifurcation analysis in weakly-coupled inductive power transfer systems. IEEE Trans. Circuits Syst. I: Regul. Pap., 63 (2016), 727738.CrossRefGoogle Scholar
[9] Wu, J.; Wang, B.; Yerazunis, W.S.; Teo, K.H.: Wireless power transfer with artificial magnetic conductors, in IEEE Wireless Power Transfer Conf. (WPTC), Perugia, 2013.Google Scholar
[10] Salas, M.; Focke, O.; Herrmann, A.S.; Lang, W.: Wireless power transmission for structural health monitoring of fiber-reinforced-composite materials. IEEE Sens. J., 14 (2014), 21712176.CrossRefGoogle Scholar
[11] Dräger, T.; Mayordomo, I.; Schuster, J.: Multi-band simultaneous inductive wireless power and data transmission, in IEEE SENSORS, Valencia, 2014.Google Scholar
[12] Chen, Z.N.; Liu, G.C.; See, T.S.P.: Transmission of RF signals between MICS loop antennas in free space and implanted in the human head. IEEE Trans. Antennas Propag., 57 (2009), 18501853.CrossRefGoogle Scholar
[13] Kim, J.; Rahmat-Samii, Y.: Implanted antennas inside a human body: simulations, designs, and characterizations. IEEE Trans. Microw. Theory Tech., 52 (2004), 19341943.CrossRefGoogle Scholar
[14] Soontornpipit, P.; Furse, C.M.; Chung, Y.C.: Design of implantable microstrip antennas for communication with medical implants. IEEE Trans. Microw. Theory Tech., 52 (2004), 19441951.CrossRefGoogle Scholar
[15] Asili, M.; Green, R.; Seran, S.; Topsakal, E.: A small implantable antenna for MedRadio and ISM bands. IEEE Antennas Wireless Propag. Lett., 11 (2012), 16831685.CrossRefGoogle Scholar
[16] Hosain, M.K. et al. : Development of a compact rectenna for wireless powering of a head-mountable deep brain stimulation device. IEEE J. Translational Eng. Health Med., 2 (2014), 113.CrossRefGoogle ScholarPubMed
[17] Hosain, M.K.; Kouzani, A.Z.; Tye, S.; Mortazavi, D.: Compact stacked planar inverted-F antenna for passive deep brain stimulation implants, in Annual Int. Conf. of the IEEE in Engineering in Medicine and Biology Society (EMBC), San Diego, 2012.Google Scholar
[18] Monti, G.; Tarricone, L.; Trane, C.: Experimental characterization of a 434 MHz wireless energy link for medical applications. Prog. Electromagn. Res. C, 30 (2012), 5364.CrossRefGoogle Scholar
[19] Li, P.; Bashirullah, R.: A wireless power interface for rechargeable battery operated medical implants. IEEE Trans. Circuits Syst., 54 (2007), 912916.CrossRefGoogle Scholar
[20] Jung, K.; Kim, Y.H.; Jung Choi, E.; Jun Kim, H.; Kim, Y.J.: Wireless power transmission for implantable devices using inductive component of closed-magnetic circuit structure, in IEEE Int. Conf. on Multisensor Fusion Integration for Intelligent Systems, Seoul, 2008.Google Scholar
[21] Lee, H.M.; Park, H.; Ghovanloo, M.: A power-efficient wireless system with adaptive supply control for deep brain stimulation. IEEE J. Solid-State Circuits, 48 (2013), 22032216.CrossRefGoogle ScholarPubMed
[22] Campi, T.; Cruciani, S.; De Santis, V.; Feliziani, M.: EMF safety and thermal aspects in a pacemaker equipped with a wireless power transfer system working at low frequency. IEEE Trans. Microw. Theory Tech., 64 (2016), 375382.Google Scholar
[23] Si, P.; Hu, A.P.; Malpas, S.; Budgett, D.: A frequency control method for regulating wireless power to implantable devices. IEEE Trans. Biomed. Circuits Syst., 2 (2008), 2229.CrossRefGoogle Scholar
[24] Xue, R.-F.; Cheng, K.-W.; Je, M.: High-efficiency wireless power transfer for biomedical implants by optimal resonant load transformation. IEEE Trans. Circuits Syst. I: Regul. Pap., 60 (2013), 867874.CrossRefGoogle Scholar
[25] Qi, X.; Zhaolong, G.; Hao, W.; Jiping, H.; Zhi-Hong, M.; Mingui, S.: Batteries not included: a mat-based wireless power transfer system for implantable medical devices as a moving target. IEEE Microw. Mag., 14 (2013), 6372.Google Scholar
[26] Jow, U.M.; Ghovanloo, M.: Modeling and optimization of printed spiral coils in air, saline, and muscle tissue environments. IEEE Trans. Biomed. Circuits Syst., 45 (2009), 2122.Google Scholar
[27] Khripkov, A.; Hong, W.; Pavlov, K.: Integrated resonant structure for simultaneous wireless power transfer and data telemetry. IEEE Antennas Wireless Propag. Lett., 11 (2012), 16591662.CrossRefGoogle Scholar
[28] Yilmaz, G.; Dehollaini, C.: An efficient wireless power link for implanted biomedical devices via resonant inductive coupling, in IEEE Radio and Wireless Symp. (RWS), Santa Clara, 2012.Google Scholar
[29]Medical Device Radiocommunications Service (MedRadio), FCC, “Federal Communications Commission,” [Online]. Available: https://www.fcc.gov/general/medical-device-radiocommunications-service-medradio.Google Scholar
[30] Monti, G.; Arcuti, P.; Tarricone, L.: Resonant inductive link for remote powering of pacemakers. IEEE Trans. Microw. Theory Tech., 63 (2015), 19.CrossRefGoogle Scholar
[31] Monti, G.; De Paolis, M.V.; Tarricone, L.: Wireless power transfer link for rechargeable deep brain stimulators, in IEEE 15th Mediterranean Microwave Symp. (MMS), Lecce, 2015.Google Scholar
[32] Monti, G.; De Paolis, M.V.; Tarricone, L.: Wireless energy link for deep brain stimulation, in Eur. Microwave Conf. (EuMC), Paris, 2015.Google Scholar
[33]‘IT'IS Foundation – Database at a Glance’, [Online]. Available http://www.itis.ethz.ch/itis-for-health/tissue-properties/database/.Google Scholar
[34] Collin, R.E.: Foundations for Microwave Engineering, 2nd ed., Wiley-IEEE Press, New York, 2001.CrossRefGoogle Scholar
[35] Inagaki, N.: Theory of image impedance matching for inductively coupled power transfer systems. IEEE Trans. Microw. Theory Tech., 62 (2014), 901908.CrossRefGoogle Scholar
[36] Fu, M.; Ma, C.; Zhu, X.: A cascaded Boost–Buck converter for high-efficiency wireless power transfer systems. IEEE Trans. Ind. Inform., 10 (2014), 19721980.CrossRefGoogle Scholar
[37] Wei, M.D.; Chang, Y.T.; Wang, D.; Tseng, C.H. and Negra, R.: Balanced RF rectifier for energy recovery with minimized input impedance variation. IEEE Trans. Microwave Theory Tech., 65 (2017), 15981604.CrossRefGoogle Scholar
[38] Kotani, K.; Sasaki, A. and Ito, T.: High-efficiency differential-drive CMOS rectifier for UHF RFIDs. IEEE J. Solid-State Circuits, 44 (2009), 30113018.CrossRefGoogle Scholar
[39] Mahmoud, M.: Efficiency improvement of differential drive rectifier for wireless power transfer applications, in 7th Int. Conf. on Intelligent Systems, Modelling and Simulation (ISMS), Bangkok, 2016.Google Scholar
[40] Cavalheiro, D.; Moll, F.; Valtchev, S.: A comparison of TFET rectifiers for RF energy harvesting applications, in 2016 IEEE Int. Power Electronics and Motion Control Conf. (PEMC), Varna, 2016.Google Scholar
[41]“IEEE Standard for Safety Levels with Respect to Human Exposure to Radiofrequency Electromagnetic Fields, 3 kHz to 300 GHz”, in IEEE Standard C95.1. 2005, 2005.Google Scholar
[42] Guideline, ICNIRP: Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz). Health Phys., 74 (1998), 494522.Google Scholar

Send article to Kindle

To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Inductive link for power and data transfer to a medical implant
Available formats
×

Send article to Dropbox

To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

Inductive link for power and data transfer to a medical implant
Available formats
×

Send article to Google Drive

To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

Inductive link for power and data transfer to a medical implant
Available formats
×
×

Reply to: Submit a response

Please enter your response.

Your details

Please enter a valid email address.

Conflicting interests

Do you have any conflicting interests? *