Hostname: page-component-99c86f546-4k54s Total loading time: 0.178 Render date: 2021-12-04T02:25:27.294Z Has data issue: true Feature Flags: { "shouldUseShareProductTool": true, "shouldUseHypothesis": true, "isUnsiloEnabled": true, "metricsAbstractViews": false, "figures": true, "newCiteModal": false, "newCitedByModal": true, "newEcommerce": true, "newUsageEvents": true }

Commissioning a new CT simulator I: CT simulator hardware

Published online by Cambridge University Press:  01 September 2007

D. Kearns*
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast, N. Ireland
M. McJury
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast, N. Ireland
*
Correspondence to: D. Kearns, Department of Medical Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Lisburn Road, Belfast BT9 7AB, N. Ireland. E-mail: denise.kearns@mpa.n-i.nhs.uk

Abstract

This paper reports on the commissioning tests performed on a new GE Lightspeed RT wide-bore computed tomography (CT) scanner with GE Advantage Sim software. This paper focuses on the laser marking system, CT hardware and the interfaces between each component of the system, and generalises the findings to most CT simulation systems currently available. A discussion on the commissioning of the virtual simulator software will follow in a separate paper. Three phantoms were used (two constructed in-house) to assist with a range of tests covering aspects such as the laser patient marking system, CT hardware, and connections between the CT scanner, virtual simulator system and the treatment planning system (TPS) and VARiS. Tests performed showed the CT simulation system to be working within acceptable tolerances suggested in the literature, and baseline data have been obtained against which future comparisons of system performance have been made. Where no tolerances were available, we have suggested suitable values. While considering tolerances on Hounsfield number variation that may lead to a dose error in excess of 2%, we found that in the case of low-kV CT scanning the range of Hounsfield numbers for dense bone was outside the acceptable limits for potential dose errors and so users were advised not to perform radiotherapy planning CT scans with kV below 100 kV.

Type
Original Article
Copyright
Copyright © Cambridge University Press 2007

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Sherouse, GW, Chaney, EL.The portable virtual simulator. Int J Radiation Oncol Biol Phys 1991; 21:475481.CrossRefGoogle ScholarPubMed
McJury, M, Dyker, K, Nakielny, R, Conway, J, Robinson, MH.Optimizing localization accuracy in head and neck, and brain radiotherapy. Br J Radiol 2006; 79:672680.CrossRefGoogle ScholarPubMed
McGee, KP, Das, IJ. Commissioning, acceptance testing, and quality assurance of a CT simulator. In: Coia, LR, Schultheiss, TE, Hanks, GE (eds). A Practical Guide to CT Simulation (pp. 5–23). Madison, Wisconsin: Advanced Medical Publishing, 1995.Google Scholar
Mutic, S, Palta, JR, Butker, EK, et al. Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66. Med Phys 2003; 30 (10):27622792.CrossRefGoogle ScholarPubMed
Aird, E, Conway, J.CT simulation for radiotherapy treatment planning. Br J Radiol 2002; 75:113.CrossRefGoogle ScholarPubMed
Conway, J, Robinson, MH. CT virtual simulation. Br J Radiol. 1997; 70:S106S118.CrossRefGoogle ScholarPubMed
Craig, T, Brochu, D, Van Dyk, J. A quality assurance phantom for three-dimensional radiation treatment planning. Int J Radiation Oncol Biol Phys. 1999; 44:955966.CrossRefGoogle ScholarPubMed
McGee, KP, Das, IJ, Sims, C.Evaluation of digitally reconstructed radiographs (DRRs) used for clinical radiotherapy: A phantom study. Med Phys 1995; 22:18151827.CrossRefGoogle ScholarPubMed
IPEM Report 91. Recommended Standards for the Routine Performance Testing of Diagnostic X-ray Imaging Systems. 2005.Google Scholar
Kilby, W, Sage, J, Rabett, V. Tolerance levels for quality assurance of electron density values generated from CT in radiotherapy treatment planning. Phys Med Biol 2002; 47:14851492.CrossRefGoogle ScholarPubMed
Knoos, T, Nilsson, M, Ahlgren, L.A method for conversion of Hounsfield number to electron density and prediction of macroscopic pair production cross-sections. Rad Oncol 1986; 5:337345.CrossRefGoogle ScholarPubMed
3
Cited by

Send article to Kindle

To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

Commissioning a new CT simulator I: CT simulator hardware
Available formats
×

Send article to Dropbox

To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

Commissioning a new CT simulator I: CT simulator hardware
Available formats
×

Send article to Google Drive

To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

Commissioning a new CT simulator I: CT simulator hardware
Available formats
×
×

Reply to: Submit a response

Please enter your response.

Your details

Please enter a valid email address.

Conflicting interests

Do you have any conflicting interests? *