Hostname: page-component-77c89778f8-vpsfw Total loading time: 0 Render date: 2024-07-16T12:32:16.712Z Has data issue: false hasContentIssue false

Direct Quantative Determination of Silica by X-ray Diffraction on Pvc Membrane Filters

Published online by Cambridge University Press:  06 March 2019

W. W. Henslee
Affiliation:
Central Laboratory, Dow Chemical U.S.A. Freeport, Texas 77541
R. E. Guerra
Affiliation:
Central Laboratory, Dow Chemical U.S.A. Freeport, Texas 77541
Get access

Abstract

Exposure of industrial workers to respirable crystalline silica is receiving considerable attention at present from industrial laboratories and regulatory agencies. Analytical methodology for rapid and accurate determinations of microgram quantities of silica is necessary. The X-ray diffraction method which has been recommended by NIOSH requires lengthy sample preparation.

An X-ray diffraction method has been developed for the quantitative analysis of α-quartz directly on polyvinyl chloride filters in the microgram range. The minimum detectable concentration of α-quartz is approximately 2μg/cm2. This represents a sensitivity comparable to that claimed using the Bumsted method. The necessity of ashing the filter and depositing the residue on silver membrane filters, as well as the addition of a standard, is avoided. This minimizes handling and alteration of the sample.

Type
X-Ray Powder Diffraction
Copyright
Copyright © International Centre for Diffraction Data 1976

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Williams, C. J. and Hawley, R. E., “An Industrial Hazard Silica Dust”, Amer. Laboratory, 17-27 (July, 1975).Google Scholar
2. NIOSH Criteria Document, “Recommendations for an Occupational Exposure Standard for Crystalline Silica”, Federal Register, Vol. 39) No. 250 (December, 1974).Google Scholar
3. Ibid, Appendix II.Google Scholar
4. Bumsted, H. E., “Determination of alpha-Quartz in the Respirable Portion of Airborne Particulates by X-ray Diffraction”, Amer. Ind. Hyg. Assoc. U., p. 150158 (April, 1973).Google Scholar
5. Edmonds, J. W., Dow Chemical Company, private communication (1975).Google Scholar
6. Pollack, S. S., “X-Ray Diffraction from Micrograms of Quartz on Filters Using a Rotating Anode”, Amer. Ind. Hyg. Assoc. J., 73-75 (January, 1975).Google Scholar
7. Gebhardt, R. F., “Respirable Dust Regulations and Measurement”, paper presented to the Sept. 1975 meeting of the Portland Cement Assoc, at Tuscon, Arizona. Google Scholar
8. McGowan, G., Dow Chemical Company, private communication (1975).Google Scholar
9. Leroux, J. and Powers, C. A., “Direct X-ray Diffraction Quantitative Analysis of Quartz in Industrial Dust Films Deposited on Silver Membrane Filters”, Staub-Relnha1t Luft, Vol. 29, No. 5, 2631 (1969).Google Scholar
10. Knight, G., Stefanich, W., and Ireland, G., “Laboratory Callbration of a Technique for Determination of Respirable Quartz in Mine Air”, Amer. Ind. Hyg. Assoc. J., 469-475 (July, 1972).Google Scholar
11. Jenkins, R. and Paolini, F. R., “An Automatic Divergence Slit for the Powder Diffractometer”, Norelco Reporter, Vol. 21, No. 1, 914 (1974).Google Scholar
12. Bradley, A. A., “The Determination of Quartz in Small Samples by an X-Ray Technique”, J. Sci. Instru., Vol. 44, 287288 (1967).Google Scholar