We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To send content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about sending content to .
To send content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about sending to your Kindle.
Note you can select to send to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
To compare the effectiveness of manual and automated methods for cleaning laparoscopic instruments.
DESIGN
Experimental laboratory study.
METHODS
We evaluated 4 methods of cleaning laparoscopic instruments: (1) manual-only cleaning and rinsing with potable tap water; (2) manual cleaning and rinsing with potable tap water, followed by ultrasonic cleaning without rinsing; (3) manual cleaning and rinsing with potable tap water followed by ultrasonic cleaning and rinsing with potable tap water; and (4) manual cleaning and rinsing with potable tap water, followed by ultrasonic cleaning and rinsing: first with potable tap water and then with sterile distilled water. Organic residues of protein, hemoglobin, and carbohydrates were evaluated using spectrophotometry.
RESULTS
The various cleaning methods tested did not result in statistically significant differences (P>.05) in the levels of investigated organic residues.
CONCLUSIONS
All cleaning and rinsing methods tested were found to be effective in reducing the levels of organic residues on laparoscopic instruments. Thus, there is no advantage gained by supplementing manual-only cleaning with automated ultrasonic cleaning, nor was there a difference between rinsing with potable tap versus sterile distilled water.
Infect Control Hosp Epidemiol 2018;39:58–63