We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Stereotactic radiosurgery (SRS) has proven itself as an effective tool in the treatment of intracranial lesions. Image-guided high dose single fraction treatments have the potential to deliver ablative doses to tumours; however, treatment times can be long. Flattening filter free (FFF) beams are available on most modern linacs and offer a higher dose rate compared to conventional flattened beams which should reduce treatment times. This study aimed to compare 6 MV FFF and 10 MV FFF to a 6 MV flattened beam for single fraction dynamic conformal arc SRS for a Varian Truebeam linac.
Materials and methods:
In total, 21 individual clinical treatment plans for 21 brain metastases treated with 6 MV were retrospectively replanned using both 6 MV FFF and 10 MV FFF. Plan quality and efficiency metrics were evaluated by analysing dose coverage, dose conformity, dose gradients, dose to normal brain, beam-on-time (BOT), treatment time and monitor units.
Results:
FFF resulted in a significant reduction in median BOT for both 6 MV FFF (57·9%; p < 0·001) and 10 MV FFF (76·3%; p < 0·001) which led to reductions in treatment times of 16·8 and 21·5% respectively. However, 6 MV FFF showed superior normal brain dose sparing (p < 0·001) and dose gradient (p < 0·001) compared to 10 MV FFF. No differences were observed for conformity.
Conclusion:
6 MV FFF offers a significant reduction in average treatment time compared to 6 MV (3·7 minutes; p = 0·002) while maintaining plan quality.