Extract

The scope of this article is to describe a transmission X-ray microscope, possible biological applications of soft X-ray microscopy and preliminary results.

For soft X-ray microscopy of biological specimens the wavelength range of 1–10 nm is best suited. Microscopy in this wavelength range requires intense X-ray sources as well as high-resolution X-ray lenses. Intense X-radiation is provided by the synchrotron radiation of electron and positron storage rings. Suited X-ray lenses are zone plates.

A theoretical treatment of the contrast mechanism and the radiation damage as well as first experiments yield the following results. Firstly, relatively thick (1–10 μm) biological specimens can be investigated. This means that unsectioned dried and even wet cells and cell organelles in a natural state can be examined. Second, it will be possible to resolve cellular aggregates in live cells with a resolution in the region of ≤10nm.

Copyright

References

Hide All
Engström, A. (1946). Quantitative micro-and histochemical elementary analysis by Roentgen absorption spectrography. Acta radiol., Suppl. 63, 1106.
Flanders, D. C., Shaver, D. C., Hawryluk, A. M.& Smith, H. I. (1980). Progress in the fabrication of artificial microstructures for X-ray imaging and spectroscopy. In Ultrasoft X-ray microscopy: Its Application to Biological and Physical Sciences. Ann. N.Y. Acad. Sci. 342, 203212.
Kirz, J. (1980). Specimen damage considerations in biological microprobe analysis. In Scanning Electron Microscopy. Chicago: SEM Inc., AMF O'Hare.
Niemann, B., Rudolph, D.& Schmahl, G. (1974). Soft X-ray imaging zone plates with large zone numbers for microscopic and spectroscopic applications. Opt. Commun. 12, 160163.
Niemann, B., Rudolph, D.& Schmahl, G. (1976). X-ray microscopy with synchrotron radiation. Appl. Optics 15, 1883.
Parsons, D. F., Matricardi, V. R., Moretz, R. C.& Turner, J. N. (1974). Electron microscopy and diffraction of wet unstained and unfixed biological objects. Adv. biol. med. Phys. 15, 161270.
Rosser, R. (1980). Proposal for a soft X-ray microscope at SSR Daresbury. Report, Blackett Laboratory, Imperial College, London, SW72BZ.
Rudolph, D.& Schmahl, G. (1980). High power zone plates for a soft X-ray microscope. In Ultrasoft X-ray microscopy: Its application to biological and physical sciences. Ann. N. Y. Acad. Sci. 342, 94104.
Sayre, D., Kirz, J., Feder, R., Kim, D. M.& Spiller, E. (1977). Transmission microscopy of unmodified biological materials: comparative radiation dosages with electrons and ultrasoft X-ray photons. Ultra- microscopy 2, 337349.
Schmahl, G., Rudolph, D.& Niemann, B. (1980). Imaging and scanning soft X-ray microscopy with zone plates. Proc. Int. Symp. ‘Scanned Image Microscopy’. The Royal Society, London.
Schmahl, G., Rudolph, D., Niemann, B.& Christ, O. (1980). X-ray microscopy of biological specimens with a zone-plate microscope. In Ultrasoft X-ray Microscopy: Its Application to Biological and Physical Sciences. Ann. N.Y. Acad. Sci. 342, 368386.

Zone-plate X-ray microscopy

  • G. Schmahl (a1), D. Rudolph (a1), B. Niemann (a1) and O. Christ (a1)

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed.