Hostname: page-component-8448b6f56d-gtxcr Total loading time: 0 Render date: 2024-04-25T06:01:41.860Z Has data issue: false hasContentIssue false

Fractured Steinmann Threaded Fixation Pins Following Vertebral Body Resection and Stabilization: Medical Device Pathological Analysis of Two Cases

Published online by Cambridge University Press:  22 February 2011

Nir Kossovsky
Affiliation:
Department of Pathology, New York Hospital, Cornell Medical Center, New York, New York
Manuel Dujovny
Affiliation:
Department of Neurological Surgery, Henry Ford Hospital, Detroit, Michigan
Dennis McDonnell
Affiliation:
Department of Neurosurgery, Medical College of Georgia, Augusta, Georgia
Ram Kossowsky
Affiliation:
Pennsylvania State University, Applied Research Laboratory, State College, Pennsylvania
Get access

Abstract

Vertebral body resection in the treatmenL of metastatic neoplasms of the spine is increasingly common for radioresistant tumors. It is also favored for treating progressive neurological disease due to tumor-associated spinal cord compression. Spine stabilization following resection is accomplished by several techniques. Increasing in popularity is the combination of threaded fixation pins and methylmethacrylate filler.

In the two cases reported in this communication, failure of threaded AISI 316 stainless steel pins occured through corrosion fatigue. Corrosion begins at the base of the threads and is exacerbated by galvanic cell formation at the methyl methacrylate-bone interface. Fatigue is experienced because the cervical and vertebral bodies are each a part of a joint that normally experiences 10 – 20 degrees of both antero-posterior and lateral flexion.

We conclude that in active patients, non-threaded cobalt chrome pins are the preferred material for stabilization of the lumbar and cervical spine following vertebral body resections.

Type
Research Article
Copyright
Copyright © Materials Research Society 1986

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

REFERENCES

1. Sundaresan, N., Galicich, J.H., Bains, M.S. and Lane, J., presented at the 34th Annual Meeting, College of Neurological Surgeons, New York, New York, 1984 (unpublished).Google Scholar
2. Harrington, K.D., J Neurosurg 61, 107 (1984).CrossRefGoogle Scholar
3. Bohlman, H.H. and Eismont, F.J., Clin Orthop 154, 57 (1981).Google Scholar
4. Bradford, D.S., Ganjavian, S., Antonious, D. et al, J Bone Joint Surg, 64A, 680 (1982).Google Scholar
5. Sundaresan, N., Galicich, J.H., Lane, J.M., J Neurosurg 60, 82 (1984).Google Scholar
6. Harrington, K.D., Sim, F.H., En-is, J.E., et al, J Bone Joint Surg, 68A, 1047 (1976).Google Scholar
7. Sundaresan, N., Galicich, J.H., Rains, M.S., et al, Cancer 53, 1393 (1984).Google Scholar
8. Boland, P.J., Lane, J.M., and Sundaresan, N., Clin Orthop 169, (1982).Google Scholar
9. Kossowsky, R., Kossovsky, N., and Dujovny, M., Med Dev Diagn Indust 7, 33 (1985).Google Scholar
10. Sugar, O., Theyear Book of Neurology and Neurosurry, edited by DeJong, R. N., Currier, R.D., and Sugar, O. (Yearbook Medical, Chicago, 1985), p 403.Google Scholar
11. Scoville, W.B., Palmer, P.H., J Neurosurg 27, 274 (1967).Google Scholar
12. Harrington, K.D., J Bone Joint Surg 53A, 36 (1981).Google Scholar
13. Green, S.A., Complications of external skeletal fixation: Causes, Prevention and Treatment, (Thomas, Springfield, 1981).Google Scholar
14. Frankel, U.H. and Nordin, M., Basic Biomechanics of the Skeletal System, (Lea G. Febiger, Philadelphia, 1980).Google Scholar
15. Foley, M.J., Calenoff, D., Hendrix, R.W., Shafer, M.F., Am J Nuentgenology 141, 373 (1983).Google Scholar
16. Scales, J.T. and Perry, P., Acta Orthopedic Belg 40, 836 (1974).Google Scholar
17. Zand, N.S., Goldstein, S.A., and Mathews, I.S., J Bone Mechanics 16, 305 (1983).Google Scholar
18. Wheeler, K.R. and James, C.A., J Biomed Mater Res 5, 267 (1971).Google Scholar
19. Dujovny, M., Kossovsky, N., Kossowsky, R., Haines, S., and Maroon, J.C., J Neurosurg 50, 368 (1979).Google Scholar
20. Kossowsky, R. (private communication).Google Scholar