The comparison between 6 MV Primus LINAC simulation output using EGSnrc and commissioning data

  • Mohammad Davoudi (a1), Ali Shabestani Monfared (a2) and Mohammad Rahgoshay (a3)
  • Please note a correction has been issued for this article.

Abstract

Introduction

Monte Carlo calculation method is considered to be the most accurate method for dose calculation in radiotherapy. The purpose of this research is comparison between 6 MV Primus LINAC simulation output with commissioning data using EGSnrc and build a Monte Carlo geometry of 6 MV Primus LINAC as realistically as possible. The BEAMnrc and DOSXYZnrc (EGSnrc package) Monte Carlo model of the LINAC head was used as a benchmark.

Methods

In the first part, the BEAMnrc was used for the designing of the LINAC treatment head. In the second part, dose calculation and for the design of 3D dose file were produced by DOSXYZnrc. The simulated PDD and beam profile obtained were compared with that calculated using commissioning data. Good agreement was found between calculated PDD (1·1%) and beam profile using Monte Carlo simulation and commissioning data. After validation, TPR20,10, TMR and Sp values were calculated in five different field.

Results

Good agreement was found between calculated values by using Monte Carlo simulation and commissioning data. Average differences for five field sizes in this approach is about 0·83% for Sp. for TPR20,10 differences for field sizes 10×10 cm2 is 0·29% and for TMR in five field sizes, the average value is ~1·6%.

Conclusion

In conclusion, the BEAMnrc and DOSXYZnrc codes package have very good accuracy in calculating dose distribution for 6 MV photon beam and it can be considered as a promising method for patient dose calculations and also the Monte Carlo model of primus linear accelerator built in this study can be used as method to calculate the dose distribution for cancer patients.

Copyright

Corresponding author

Correspondence to: Ali Shabestani Monfared Cell, Cancer Research Center, Medical Physics Department, Rajaee Oncology Hospital, Babol University of Medical Sciences, Babol, Iran Tel: 98 9111230475, E-mail: monfared1345@gmail.com

References

Hide All
1. Toossi, M B, Momennezhad, M, Hashemi, M. Monte Carlo simulation of a linear accelerator and electron beam parameters used in radiotherapy. IJMP 2009; 6 (2): 1118.
2. Serrano, B, Hachem, A, Franchisseu, E et al. Monte Carlo simulation of a medical linear accelerator for radiotherapy use. Radiat Prot Dosimetry 2006; 119 (1-4): 506509.
3. Khan, F M, Gibbons, J. P. The Physics of Radiation Therapy, 4th edition. Philadelphia, PA: Lippincott Williams & Wilkins, 2014: 153154.
4. Podgorsak, E B. Radiation Oncology Physics: A Handbook for Teachers and Students. IAEA, Vienna, Austria, 2005.
5. IAEA. Absorbed Dose Determination in External Beam Radiotherapy (TRS 398)12. IAEA, Vienna, Austria, 2000: 68–69.
6. Pena, J, Franco, L, Gomez, F et al. Commissioning of a medical accelerator photon beam Monte Carlo simulation using wide-field profiles. Phys Med Biol 2004; 49: 49294942.
7. Aljamal, M, Zakaria, A. Monte Carlo modeling of a Siemens Primus 6 MV photon beam linear accelerator. Aust J Basic Appl Sci 2013; 7 (10): 340346.
8. Jabbari, K, Sberi, H, Tvakoli, M, Amouheydari, A. Monte Carlo simulation of Siemens ONCOR linear accelerator with BEAMnrc and DOSXYZnrc code. J Med Signals Sens 2013; 3 (3): 172174.
9. Stathakis, S, Balbi, F, Chronopoulos, A, Papanikolaou, N. Monte Carlo modeling of linear accelerator using distributed computing. J BUON 2016; 21 (1): 252260.
10. McKerracher, C, Thwaites, D. Phantom scatter factors for small MV photon fields. Radiother Oncol 2008; 86: 272275.
11. Chang, K P, Wei Wang, Z, Shiau, A. Determining optimization of the initial parameters in Monte Carlo simulation for linear accelerator radiotherapy. Radiation Physics and Chemistry: Elsevier, 2014; 95: 161165.
12. Toutaoui, A, Ait chikh, S, Khelassi-Toutaoui, N, Hattali, B. Monte Carlo photon beam modeling and commissioning for radiotherapy dose calculation algorithm. Physica Medica: Elsevier, 2014: 15.
13. Ding, G X. Using Monte Carlo simulations to commission photon beam output factors-a feasibility study. Phys Med Biol 2003; 48: 38653874.
14. Verhaegen, F., Seuntjens, J. Monte Carlo modelling of external radiotherapy photon beams. Phys Med Biol 2003; 48: R107R164.

Keywords

Related content

Powered by UNSILO

The comparison between 6 MV Primus LINAC simulation output using EGSnrc and commissioning data

  • Mohammad Davoudi (a1), Ali Shabestani Monfared (a2) and Mohammad Rahgoshay (a3)
  • Please note a correction has been issued for this article.

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed.

A correction has been issued for this article: