Hostname: page-component-684bc48f8b-2l47r Total loading time: 0.348 Render date: 2021-04-11T13:59:18.447Z Has data issue: true Feature Flags: { "shouldUseShareProductTool": true, "shouldUseHypothesis": true, "isUnsiloEnabled": true, "metricsAbstractViews": false, "figures": false, "newCiteModal": false, "newCitedByModal": true }

The effect of leg position on the dose distribution of intracavitary brachytherapy for cervical cancer: 3D computerised tomography plan evaluation and in vivo dosimetric study

Published online by Cambridge University Press:  20 May 2016

Mustafa Cengiz
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Fatma Colak
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Demet Yildiz
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Ali Dogan
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Gokhan Ozyigit
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Ferah Yildiz
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Murat Gurkaynak
Affiliation:
Department of Radiation Oncology, Hacettepe University, Ankara, Turkey
Corresponding
E-mail address:

Abstract

Purpose

To evaluate the impact of leg position on the dose distribution during intracavitary brachytherapy for cervical cancer.

Patients and methods

This prospective study was performed on 11 women with cervical cancer who underwent intracavitary brachytherapy. After insertion of the brachytherapy applicator, two sets of computed tomography slices were taken including pelvis, one with straight leg and one with leg flexion position with knee support. The dose (7 Gy) was prescribed to point A. The radiotherapy plan was run on the Plato Planning Software System V14·1 to get the dose distributions. Also, rectum and bladder doses were measured for both leg positions during the treatment. The doses and volumes of organs were compared via the Wilcoxon signed-rank test by using Statistical Package for the Social Sciences 11·5 statistical software.

Results

No significant difference regarding the dose distributions and volumes of target, sigmoid and bladder due to leg position was observed, either on 3D planning or on in vivo dose measurements. However, there were significant differences for 25 and 50% isodose coverage of rectum in favour of straight leg position (p=0·026). There were no significant differences regarding maximum doses in any critical organ.

Conclusion

Difference in leg position caused only a small change in rectum dose distribution and did not cause any other change in either dose distributions or in vivo measured doses of both target and critical organs during cervical brachytherapy. Straight leg position appears better with regard to rectum dose.

Type
Original Articles
Copyright
© Cambridge University Press 2016 

Access options

Get access to the full version of this content by using one of the access options below.

References

1. Atahan, IL, Yildiz, F, Ozyar, E et al. Radiotherapy in the adjuvant setting of cervical carcinoma: treatment, results, and prognostic factors. Int J Gynecol Cancer 2007; 17: 813820.CrossRefGoogle ScholarPubMed
2. Okkan, S, Atkovar, G, Sahinler, I et al. Results and complications of high dose rate and low dose rate brachytherapy in carcinoma of the cervix: cerrahpasa experience. Radiother Oncol 2003; 67: 97105.CrossRefGoogle ScholarPubMed
3. Weeks, KJ, Montana, GS. Three-dimensional applicator system for carcinoma of the uterine cervix. Int J Radiat Oncol Biol Phys 1997; 37: 455463.CrossRefGoogle ScholarPubMed
4. Cengiz, M, Gurdalli, S, Selek, U et al. Effect of bladder distension on dose distribution of intracavitary brachytherapy for cervical cancer: three-dimensional computed tomography plan evaluation. Int J Radiat Oncol Biol Phys 2008; 70: 464468.CrossRefGoogle ScholarPubMed
5. Schoeppeel, SL, La Vigne, ML, Martel, MK et al. Three-dimensional treatment planning of intracavitary gynecologic implants: analysis of ten cases and implications for dose specification. Int J Radiat Oncol Biol Phys 1993; 28: 277283.CrossRefGoogle Scholar
6. Cengiz, M, Selek, U, Genc, M et al. Comment on ‘Correlation between the treated volume, the GTV and the CTV at the time of brachytherapy and histopathologic findings in 33 patients with operable cervix carcinoma’. Radiother Oncol 2005; 75: 367368.CrossRefGoogle Scholar
7. Pelloski, CE, Palmer, M, Chronowski, GM et al. Comparison between CT-based volumetric calculations and ICRU reference-point estimates of radiation dose delivered to bladder and rectum during intracavitary radiotherapy for cervical cancer. Int J Radiat Oncol Biol Phys 2005; 62: 131137.CrossRefGoogle ScholarPubMed
8. Nag, S, Gardenes, H, Chang, S et al. Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: report from image-guided bracyhtherapy working group. Int J Radiat Oncol Biol Phys 2004; 60: 11601172.CrossRefGoogle ScholarPubMed
9. Pötter, R, Haie-Meder, C, Limbergen, E et al. Recommendations from gynecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image-based treatment planning in cervix cancer bracyhterapy-3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol 2006; 78: 6777.CrossRefGoogle Scholar
10. Steenbakkers, RJ, Duppen, JC, Betgen, A et al. Impact of knee support and shape of table top on rectum and prostate position. Int J Radiat Oncol Biol Phys 2004; 60: 13641372.CrossRefGoogle Scholar
11. Nag, S, Erickson, B, Thomadsen, B et al. The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. Int J Radiat Oncol Biol Phys 2000; 48: 201211.CrossRefGoogle ScholarPubMed
12. Chao, KS, Cengiz, M, Herzog, T. Gynocologic tumors. In Gregoire V, Scalliet P, Ang KK (eds) Clinical Target Volumes Conformal and Intensity Modulated Radiation Therapy. Berlin: Springer, 2004; pp. 171186.CrossRefGoogle Scholar
13. Schoeppel, SL, LaVigne, ML, Martel, MK et al. Three-dimensional treatment planning of intracavitary gynecologic implants: analysis of ten cases and implications for dose specification. Int J Radiat Oncol Biol Phys 1994; 28: 277283.CrossRefGoogle ScholarPubMed
14. Thomadsen, BR, Shahabi, S, Stitt, JA et al. High dose rate brachytherapy for carcinoma of the cervix: the Madison system-II – procedural and physical considerations. Int J Radiat Oncol Biol Phys 1992; 24: 349357.CrossRefGoogle ScholarPubMed
15. Pham, HT, Chen, Y, Rouby, E et al. Changes in high dose rate tandem and ovoid applicator positions during treatment in an unfixed brachytherapy system. Radiology 1998; 206: 525531.CrossRefGoogle Scholar
16. Cheng, JC, Peng, LC, Chen, YH et al. Unique role of proximal rectal dose in late rectal complications for patients with cervical cancer undergoing high-dose-rate intracavitary brachytherapy. Int J Radiat Oncol Biol Phys 2003; 57: 10101018.CrossRefGoogle ScholarPubMed
17. Wang, X, Liu, R, Yank, K et al. High dose rate versus low dose rate intracavity brachytherapy for locally advanced uterine cervix cancer. Cochrane Database Syst Rev 2010; 7: CD 007563.Google Scholar
18. Usla, CW, Wabersieb, A, Potter, R, Georg, D. In-vivo dosimetry for gynaecological brachytherapy: physical and clinical considerations. Radiother Oncol 2005; 77: 310317.Google Scholar

Full text views

Full text views reflects PDF downloads, PDFs sent to Google Drive, Dropbox and Kindle and HTML full text views.

Total number of HTML views: 5
Total number of PDF views: 43 *
View data table for this chart

* Views captured on Cambridge Core between September 2016 - 11th April 2021. This data will be updated every 24 hours.

Send article to Kindle

To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

Find out more about the Kindle Personal Document Service.

The effect of leg position on the dose distribution of intracavitary brachytherapy for cervical cancer: 3D computerised tomography plan evaluation and in vivo dosimetric study
Available formats
×

Send article to Dropbox

To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

The effect of leg position on the dose distribution of intracavitary brachytherapy for cervical cancer: 3D computerised tomography plan evaluation and in vivo dosimetric study
Available formats
×

Send article to Google Drive

To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

The effect of leg position on the dose distribution of intracavitary brachytherapy for cervical cancer: 3D computerised tomography plan evaluation and in vivo dosimetric study
Available formats
×
×

Reply to: Submit a response


Your details


Conflicting interests

Do you have any conflicting interests? *