        To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

        Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

        Find out more about the Kindle Personal Document Service.

        A Procedure for Testing Polycrystalline Ice in Uniaxial Tension
        Available formats
        ×

        Send article to Dropbox

        To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

        A Procedure for Testing Polycrystalline Ice in Uniaxial Tension
        Available formats
        ×

        Send article to Google Drive

        To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

        A Procedure for Testing Polycrystalline Ice in Uniaxial Tension
        Available formats
        ×
Export citation

Abstract

A procedure is described for testing ice in tension. It incorporates a one-piece lucite mold, two carpeted “Synthane” end-caps and a pair of yokes anti ball joints. Use of the procedure has resulted in a successful test rate greater than 80%.

This note outlines a procedure, developed and successfully employed over the past three years, to test polycrystalline ice in uniaxial tension. Details beyond those presented here may be obtained from Currier (unpublished), Currier and others (1983) and from Lim (unpublished); applications and test results at T = −20 to −5°C at 10−6 s−1 are described by Currier and Schulson (1982) and by Schulson and others (1984).

The brittle nature of ice when tested in tension requires (1) a capability to consistently produce accurately dimensioned samples and (2) a method of applying a pure uniaxial tensile stress.

The method used to prepare samples is similar to that developed by Cole (1979). A cylindrical mold is charged with seed grains, evacuated, flooded with degassed and deionized water, and frozen radially inward. During freezing the system remains open to water flow to flush impurities and to avoid the development of internal stresses due to the expansion of freezing water.

The mold (Fig. 1) consists of a one-piece lucite (polymethyl-methacrylate) cylinder into which fit lucite alignment caps. Right circular cylinders of ice, 9.1 cm in diameter and 23.1 cm in length, are extracted by warming the cylinder surface gently with a warm-air gun and by pushing by hand. Careful machining insures that the inside diameter of the cylinder is uniform to ± 0.0013 cm, that the ends are parallel and perpendicular to the wall and that the O-ring bearing surfaces on each alignment cap are parallel.

Fig. 1. One-piece lucite cylinder with alignment caps. Total cylinder length is 35 Cm.

To apply the tensile load, linen-based phenolic (Synthane) end-caps, as proposed by Mellor (Schwarz and others, 1981, p. 250), are attached to each alignment cap (Fig. 2). Upon freezing, they become integral parts of the sample. The end-caps are 9.1 cm in diameter (the sample diameter) and 3.2 cm in thickness. A threaded hole, 2.54 cm in diameter, in the center of each end-cap allows attachment to the testing machine.

Fig. 2. Jute-backed carpet epoxied to Synthane endcaps.

Initially the ice was frozen directly to end-caps whose bonding surfaces had been roughened by passing a single-point threading tool across the surface as the cap was rotated in a lathe. While this method provided a sufficiently strong bond between ice and endcaps, the fracture surface was typically “domed” and, although entirely within the ice, very near an endcap.

The domed surface indicated the presence of induced shear stresses at the ice/end-cap interface owing to differences in the elastic moduli and thermal expansion coefficients of ice and Synthane (for a further analysis see Currier and others, 1983, p. 6–8). When combined with the applied tensile stress, shear stresses, if not avoided, produce a principal stress greater than the applied stress and the sample fractures prematurely.

It was found that jute-backed carpet epoxied to each roughened end-cap surface yielded fracture surfaces normal to the tensile axis. The addition of a carpet layer provides sufficient compliance at the ice/end-cap interface to prevent the development of large shear stresses. In addition, it enhances the strength of the ice/end-cap bond. The absence of shear stresses at the end zones also prevents the need for reducing the cross-sectional area within the gage section: only right circular cylinders need be tested.

To insure the application of purely uniaxial load, a steel assembly consisting of a ball-joint and yoke was developed to connect the sample end-caps to the testing machine (Fig. 3). The ball-joints, containing a sintered iron ball which rotates freely in a carburized steel race, correct small vertical misalignments and remain threaded into the crossheads of the testing machine. A steel yoke and pin were machined for each ball-joint such that the iron ball fitted snugly into the yoke and the pin slid through to provide a rapid connection between yoke and ball-joint. In this way the ice sample becomes the “weak link” in a chain significantly more rigid than the cable system proposed by Schwarz and others (1981, p. 249–50.

Fig. 3. Yoke and ball-joint assembly. Steel connecting pin is 4.1 cm in length.

Until care was taken to control circumferential variations in end-cap thickness and overall sample dimensions, less than one half of the specimens resulted in successful tests. (A successful test results in fracture normal to the tensile axis within the central third of the sample.) However, once variations in end-cap thickness were reduced to ± 0.0013 cm and a mold was developed capable of consistently controlling sample length to ± 0.004 cm and diameter to ± 0.0013 cm, greater than 80% of the samples prepared and tested using this procedure have ended in successful tests. This experience indicates the extreme importance of sample alignment when testing ice in tension.

Acknowledgements

We wish to thank D. Cole and M. Mellor of the U.S. Army Cold Regions Research and Engineering Laboratory, Hanover, New Hampshire, for valuable assistance during the early stages of this work. Also, we acknowledge financial support from the U.S. Army Research Office, Contract No. DAAG 29–80-C-0064, and from the Mobil Corporation.

References

Cole, D.M. 1979. Preparation of polycrystalline ice specimens for laboratory experiments. Cold Regions Science and Technology, Vol. 1, No. 2, p. 15359.
Currier, J.H. Unpublished. The brittle to ductile transition in polycrystalline ice under tension. [M.Sc. thesis, Dartmouth College, 1981.]
Currier, J.H., and Schulson, E.M. 1982. The tensile strength of ice as a function of grain size. Acta Metallurgica, Vol. 30, No. 8, p. 151114.
Currier, J.H., and others. 1983. A study of the tensile strength of ice as a function of grain size, by Currier,, J.H., Schulson,, E.M. and St. Lawrence, W.F.. CRREL Report (Hanover, N.H.) 83–14.
Lim, P.N. Unpublished. The effects of temperature and grain size on the tensile strength of ice. [M.Sc. thesis, Dartmouth College, 1983.]
Schulson, E.M., and others. 1984. A brittle to ductile transition in polycrystalline ice under tension, by Schulson,, E.M., Lim,, P.N. and Lee, R.W. Philosophical Magazine A, Vol. 49, No. 3, p. 35363.
Schwarz, J., and others. 1981. Standardized testing methods for measuring mechanical properties of ice, by Schwarz, J. [and 6 others]. Cold Regions Science and Technology, Vol. 4, No. 3, p. 24553.