Outbreak of Pseudomonas aeruginosa Surgical Site Infections after Arthroscopic Procedures: Texas, 2009

  • Pritish K. Tosh (a1) (a2), Maureen Disbot (a3), Jonathan M. Duffy (a1) (a2), Marc L. Boom (a3), Gary Heseltine (a4), Arjun Srinivasan (a2), Carolyn V. Gould (a2) and Sandra I. Berríos-Torres (a2)...

Abstract

Setting.

Seven organ/space surgical site infections (SSIs) that occurred after arthroscopic procedures and were due to Pseudomonas aeruginosa of indistinguishable pulsed-field gel electrophoresis (PFGE) patterns occurred at hospital X in Texas from April 22, 2009, through May 7, 2009.

Objective.

To determine the source of the outbreak and prevent future infections.

Design.

Infection control observations and a case-control study.

Methods.

Laboratory records were reviewed for case finding. A case-control study was conducted. A case patient was defined as someone who underwent knee or shoulder arthroscopy at hospital X during the outbreak period and subsequently developed organ/space SSI due toP. aeruginosa. Cultures of environmental and surgical equipment samples were performed, and selected isolates were analyzed by PFGE. Surgical instrument reprocessing practices were reviewed, and surgical instrument lumens were inspected with a borescope after reprocessing to assess cleanliness.

Results.

The case-control study did not identify any significant patient-related or operator-related risk factors. P. aeruginosa grew from 62 of 388 environmental samples. An isolate from the gross decontamination sink had a PFGE pattern that was indistinguishable from that of the case patient isolates. All surgical instrument cultures showed no growth. Endoscopic evaluation of reprocessed arthroscopic equipment revealed retained tissue in the lumen of both the inflow/outflow cannulae and arthroscopic shaver handpiece. No additional cases occurred after changes in instrument reprocessing protocols were implemented. After this outbreak, the US Food and Drug Administration released a safety alert about the concern regarding retained tissue within arthroscopic shavers.

Conclusions.

These SSIs were likely related to surgical instrument contamination with P. aeruginosa during instrument reprocessing. Retained tissue in inflow/outflow cannulae and shaver handpieces could have allowed bacteria to survive sterilization procedures.

Copyright

Corresponding author

Mayo Clinic Division of Infectious Diseases, 200 First Street SW, Rochester, MN 55905 (tosh.pritish@mayo.edu)

Footnotes

Hide All
a

Pritish K. Tosh, MD, Mayo Clinic Division of Infectious Diseases

The findings and conclusions in this report are those of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention.

Footnotes

References

Hide All
1. Babcock, HM, Carroll, C, Matava, M, L'Ecuyer, P, Fraser, V. Surgical site infections after arthroscopy: outbreak investigation and case control study. Arthroscopy. 2003;19:172181.
2. Armstrong, RW, Bolding, F, Joseph, R. Septic arthritis following arthroscopy: clinical syndromes and analysis of risk factors. Arthroscopy. 1992;8:213223.
3. Armstrong, RW, Bolding, F. Septic arthritis after arthroscopy: the contributing roles of intraarticular steroids and environmental factors. Am J Infect Control 1994;22:1618.
4. D'Angelo, GL, Ogilvie-Harris, DJ. Septic arthritis following arthroscopy, with cost/benefit analysis of antibiotic prophylaxis. Arthroscopy. 1988;4:1014.
5. Ajemian, E, Andrews, L, Hryb, K, Klimek, JJ. Hospital-acquired infections after arthroscopic knee surgery: a probable environmental source. Am J Infect Control 1987;15:159162.
6. Blevins, FT, Salgado, J, Wascher, DC, Koster, F. Septic arthritis following arthroscopic meniscus repair: a cluster of three cases. Arthroscopy. 1999;15:3540.
7. Lopansri, B, Taylor, C, Anderson, V, Pombo, D, Burke, J. Protracted outbreak of postarthroscopy infections associated with flash sterilization of instruments. In: Program and abstracts of the Fifth Decennial International Conference on Healthcare-Associated Infections. Atlanta: Society for Healthcare Epidemiology of America, 2010. Abstract 400.
8. Parada, SA, Grassbaugh, JA, DeVine, JG, Arrington, ED. Instrumentation-specific infection after anterior cruciate ligament reconstruction. Sports Health 2009;1:481485.
9. Viola, R, Marzano, N, Vianello, R. An unusual epidemic of Staphylococcus-negative infections involving anterior cruciate ligament reconstruction with salvage of the graft and function. Arthroscopy. 2000;16:173177.
10. Johnson, LL, Shneider, DA, Austin, MD, Goodman, FG, Bullock, JM, DeBruin, JA. Two per cent glutaraldehyde: a disinfectant in arthroscopy and arthroscopic surgery. J Bone Joint Surg Am 1982;64:237239.
11. McAllister, DR, Parker, RD, Cooper, AE, Recht, MP, Abate, J. Outcomes of postoperative septic arthritis after anterior cruciate ligament reconstruction. Am J Sports Med 1999;27:562570.
12. Sherman, OH, Fox, JM, Snyder, SJ, et al. Arthroscopy— “no-problem surgery”: an analysis of complications in two thousand six hundred and forty cases. J Bone Joint Surg Am 1986;68:256265.
13. Rutala, WA, Weber, DJ; Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. http://www.cdc.gov/hicpac/Disinfection_Sterilization/toc.html.
14. US Food and Drug Administration. Arthroscopic Shavers: Ongoing Safety Review. http://www.fda.gov/Safety/MedWatch/Safetylnformation/SafetyAlertsforHumanMedicalProducts/ucml70730.htm. July 7, 2009. Accessed December 3, 2010.
15. US Food and Drug Administration. Ongoing Safety Review of Arthroscopic Shavers.http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucml70639.htm. 2009. Accessed December 3, 2010.

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed