Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination

  • Jennifer L. Cadnum (a1) (a2), Myreen E. Tomas (a1), Thriveen Sankar (a1) (a2), Annette Jencson (a1), J. Itty Mathew (a2), Sirisha Kundrapu (a2) and Curtis J. Donskey (a2) (a3)...

Abstract

OBJECTIVE

To determine the effect of variation in test methods on performance of an ultraviolet-C (UV-C) room decontamination device.

DESIGN

Laboratory evaluation.

METHODS

We compared the efficacy of 2 UV-C room decontamination devices with low pressure mercury gas bulbs. For 1 of the devices, we evaluated the effect of variation in spreading of the inoculum, carrier orientation relative to the device, type of organic load, type of carrier, height of carrier, and uninterrupted versus interrupted exposures on measured UV-C killing of methicillin-resistant Staphylococcus aureus and Clostridium difficile spores.

RESULTS

The 2 UV-C room decontamination devices achieved similar log10 colony-forming unit reductions in the pathogens with exposure times ranging from 5 to 40 minutes. On steel carriers, spreading of the inoculum over a larger surface area significantly enhanced killing of both pathogens, such that a 10-minute exposure on a 22-mm2 disk resulted in greater than 2 log reduction in C. difficile spores. Orientation of carriers in parallel rather than perpendicular with the UV-C lamps significantly enhanced killing of both pathogens. Different types of organic load also significantly affected measured organism reductions, whereas type of carrier, variation in carrier height, and interrupted exposure cycles did not.

CONCLUSIONS

Variation in test methods can significantly impact measured reductions in pathogens by UV-C devices during experimental testing. Our findings highlight the need for standardized laboratory methods for testing the efficacy of UV-C devices and for evaluations of the efficacy of short UV-C exposure times in real-world settings.

Infect Control Hosp Epidemiol 2016;37:555–560

  • View HTML
    • Send article to Kindle

      To send this article to your Kindle, first ensure no-reply@cambridge.org is added to your Approved Personal Document E-mail List under your Personal Document Settings on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part of your Kindle email address below. Find out more about sending to your Kindle. Find out more about sending to your Kindle.

      Note you can select to send to either the @free.kindle.com or @kindle.com variations. ‘@free.kindle.com’ emails are free but can only be sent to your device when it is connected to wi-fi. ‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.

      Find out more about the Kindle Personal Document Service.

      Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination
      Available formats
      ×

      Send article to Dropbox

      To send this article to your Dropbox account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Dropbox.

      Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination
      Available formats
      ×

      Send article to Google Drive

      To send this article to your Google Drive account, please select one or more formats and confirm that you agree to abide by our usage policies. If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your <service> account. Find out more about sending content to Google Drive.

      Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination
      Available formats
      ×

Copyright

Corresponding author

Address correspondence to Curtis J. Donskey, MD, Geriatric Research, Education, and Clinical Center 1110W, Cleveland VA Medical Center, 10701 East Blvd, Cleveland, OH 44106 (curtisd123@yahoo.com).

References

Hide All
1. Rutala, WA, Weber, DJ. The role of the environment in transmission of Clostridium difficile infection in health care facilities. Infect Control Hosp Epidemiol 2011;32:207209.
2. Owens, MU, Deal, DR, Shoemaker, MO, et al. High-dose ultraviolet C light inactivates spores of Bacillus subtilis var. niger and Bacillus anthracis Sterne on non-reflective surfaces. Appl Biosafety 2005;10:240247.
3. Nerandzic, MM, Cadnum, JL, Pultz, MJ, Donskey, CJ. Evaluation of an automated ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens in hospital rooms. BMC Infect Dis 2010;10:197.
4. Rutala, WA, Gergen, MF, Weber, DJ. Room decontamination with UV radiation. Infect Control Hosp Epidemiol 2010;3110:10251029.
5. Nerandzic, MM, Fisher, CW, Donskey, CJ. Sorting through the wealth of options: comparative evaluation of two ultraviolet disinfection systems. PLoS One 2014;23(9):e107444.
6. Boyce, JM, Havill, NL, Moore, BA. Terminal decontamination of patient rooms using an automated mobile UV light unit. Infect Control Hosp Epidemiol 2011;32:737742.
7. Havill, NL, Moore, BA, Boyce, JM. Comparison of the microbiological efficacy of hydrogen peroxide vapor and ultraviolet light processes for room decontamination. Infect Control Hosp Epidemiol 2012;33:507512.
8. Nerandzic, MM, Thota, P, Sankar, CT, et al. Evaluation of a pulsed xenon ultraviolet disinfection system for reduction of healthcare-associated pathogens in hospital rooms. Infect Control Hosp Epidemiol 2015;36:192197.
9. Zhang, A, Nerandzic, MM, Kundrapu, S, Donskey, CJ. Does organic material on hospital surfaces reduce the effectiveness of hypochlorite and UV radiation for disinfection of Clostridium difficile? Infect Control Hosp Epidemiol 2013;34:11061108.
10. Rutala, WA, Gergen, MF, Tande, BM, Weber, DJ. Room decontamination using an ultraviolet-C device with short ultraviolet exposure time. Infect Control Hosp Epidemiol 2014;35:10701072.
11. Jinadatha, C, Quezada, R, Huber, TW, Williams, JB, Zeber, JE, Copeland, LA. Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on contamination levels of methicillin-resistant Staphylococcus aureus . BMC Infect Dis 2014;14:187.
12. Stibich, M, Stachowiak, J, Tanner, B, et al. Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on hospital operations and microbial reduction. Infect Control Hosp Epidemiol 2011;32:286288.
13. Anderson, DJ, Gergen, MF, Smathers, E, et al. Decontamination of targeted pathogens from patient rooms using an automated ultraviolet-C-emitting device. Infect Control Hosp Epidemiol 2013;34:466471.
14. Levin, J, Riley, LS, Parrish, C, English, D, Ahn, S. The effect of portable pulsed xenon ultraviolet light after terminal cleaning on hospital-associated Clostridium difficile infection in a community hospital. Am J Infect Control 2013;41:746748.
15. Haas, JP, Menz, J, Dusza, S, Montecalvo, MA. Implementation and impact of ultraviolet environmental disinfection in an acute care setting. Am J Infect Control 2014;42:586590.
16. Simmons, S, Morgan, M, Hopkins, T, Helsabeck, K, Stachowiak, J, Stibich, M. Impact of a multi-hospital intervention utilising screening, hand hygiene education and pulsed xenon ultraviolet (PX-UV) on the rate of hospital associated methicillin resistant Staphylococcus aureus infection. J Infect Prevent 2013;14:13.
17. ASTM International. Designation E2197: standard quantitative disk carrier test method for determining bactericidal, virucidal, fungicidal, mycobactericidal, and sporicidal activities of chemicals. West Conshohocken, PA: ASTM, 2011.
18. Sorg, JA, Sonenshein, AL. Bile salts and glycine as cogerminants for Clostridium difficile spores. J Bacteriol 2008;190:25052512.
19. R Development Core Team. R: a language and environment for statistical computing. R project website. http://www.R-project.org/. Published 2012. Accessed January 9, 2016.
20. Moore, G, Ali, S, Cloutman-Green, EA, et al. Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet. BMC Infect Dis 2012;12:174.
21. Guidance for the efficacy evaluation of products with sporicidal claims against Clostridium difficile (June 2014). US Environmental Protection Agency website. http://www.epa.gov/pesticide-registration/guidance-efficacy-evaluation-products-sporicidal-claims-against-clostridium. Accessed January 1, 2015.
22. Nerandzic, MM, Donskey, CJ. Triggering germination represents a novel strategy to enhance killing of Clostridium difficile spores. PLoS One 2010;5(8):e12285.
23. Nerandzic, MM, Donskey, CJ. Activate to eradicate: inhibition of Clostridium difficile spore outgrowth by the synergistic effects of osmotic activation and nisin. PLoS One 2013;8:e54740.

Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination

  • Jennifer L. Cadnum (a1) (a2), Myreen E. Tomas (a1), Thriveen Sankar (a1) (a2), Annette Jencson (a1), J. Itty Mathew (a2), Sirisha Kundrapu (a2) and Curtis J. Donskey (a2) (a3)...

Metrics

Altmetric attention score

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed