Comparison of cardiac output measurements by arterial trans-cardiopulmonary and pulmonary arterial thermodilution with direct Fick in septic shock

  • G. Marx (a1), T. Schuerholz (a1), R. Sümpelmann (a2), T. Simon (a1) and M. Leuwer (a3)...

Abstract

Summary

Background and objective: The aim of this study was to compare cardiac output (CO) measurements by arterial trans-cardiopulmonary thermodilution (ATD) and pulmonary arterial thermodilution (PATD) with CO estimated on the basis of the Fick calculation via a metabolic monitor in septic shock.

Methods: In a prospective animal study 20 anaesthetized, ventilated pigs (20.9 ± 1.9 kg) were investigated. Septic shock was induced with faecal peritonitis. A pulmonary artery catheter was used for conventional measurement of CO. Simultaneously ATD was measured with a thermistor tipped catheter inserted into right carotid artery. Whole body oxygen consumption was measured by indirect calorimetry. Eighty data pairs of simultaneous CO measurements were analysed.

Results: CO measured with Fick and that measured with PATD were significantly correlated (r = 0.94, r2 = 0.87, P < 0.001). Mean CO measured by PATD was 94.3 ± 40.1 mL min−1 kg−1. Bias was 10.1 mL min−1 kg−1 (95% confidence interval (CI): 6.0–14.2 mL min−1 kg−1) with limits of agreement of −26.8 to 47.0 mL min−1 kg−1 . Correlation between Fick derived CO estimation and ATD CO was similar (r = 0.91, r2 = 0.83, P < 0.001). Mean CO measured by trans-cardiopulmonary thermodilution was 104.3 ± 43.2 mL min−1 kg−1 . Bias was 0.75 mL min−1 kg−1 (95% CI: −3.8 to 5.3 mL min−1 kg−1) with limits of agreement of −39.7 to 41.2 mL min−1 kg−1.

Conclusions: Even during haemodynamic instability in septic shock the correlation of arterial trans-cardiopulmonary thermodilution and PATD derived CO with direct Fick was good. As arterial trans-cardiopulmonary thermodilution is less invasive than PATD, the former may offer practical advantages.

Copyright

Corresponding author

Correspondence to: Prof. Dr. med. Gernot Marx, Klinik für Anästhesiologie und Intensivtherapie, Klinikum der Friedrich-Schiller-Universität Jena, Erlanger Allee 101, 07747 Jena, Germany. E-mail: Gernot.Marx@med.uni-jena.de; Tel: +49 36 41 932 3120; Fax: +49 36 41 932 3102/22

References

Hide All

References

Sakka S, Meier-Hellmann A. Evaluation of cardiac output and cardiac preload. In: Vincent J, ed. Yearbook of Intensive Care and Emergency Medicine.Berlin, Germany: Springer, 2000: 671679.
Perel A, Berkenstedt H, Segal E. Continuous arterial thermodilution cardiac output and derived variables. In: Vincent J, ed. Yearbook of Intensive Care and Emergency Medicine.Berlin, Germany: Springer, 1999: 459467.
Böck J, Barker W, Mackersie R, Tranbaugh R, Lewis F. Cardiac output measurement using femoral arterial artery thermodilution in patients. J Crit Care 1989; 4: 105111.
Reinhart K, Sakka SG, Meier-Hellmann A. Haemodynamic management of a patient with septic shock. Eur J Anaesth 2000; 17: 617.
Buhre W, Weyland A, Kazmaier S, et al. Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999; 13: 437440.
Goedje O, Hoeke K, Lichtwarck-Aschoff M, et al. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27: 24072412.
Hoeper MM, Maier R, Tongers J, et al. Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension. Am J Res Crit Care Med 1999; 160: 535541.
Sherman MS, Kosinski R, Paz HL, Campbell D. Measuring cardiac output in critically ill patients: disagreement between thermodilution-, calculated-, expired gas-, and oxygen consumption-based methods. Cardiology 1997; 88: 1925.
Dhingra VK, Fenwick JC, Walley KR, Chittock DR, Ronco JJ. Lack of agreement between thermodilution and fick cardiac output in critically ill patients. Chest 2002; 122: 990997.
Tibby SM, Hatherill M, Marsh MJ, et al. Clinical validation of cardiac output measurements using femoral artery thermodilution with direct Fick in ventilated children and infants. Intens Care Med 1997; 23: 987991.
Gonzalez J, Delafosse C, Fartoukh M, et al. Comparison of bedside measurement of cardiac output with the thermodilution method and the Fick method in mechanically ventilated patients. Crit Care 2003; 7: 171178.
Sandham JD, Hull RD, Brant RF, et al. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. New Engl J Med 2003; 348: 514.
Connors JrAF, Speroff T, Dawson NV, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 1996; 276: 889897.
Makita K, Nunn JF, Royston B. Evaluation of metabolic measuring instruments for use in critically ill patients. Crit Care Med 1990; 18: 638644.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307310.
Segal E, Katzenelson R, Berkenstadt H, Perel A. Transpulmonary thermodilution cardiac output measurement using the axillary artery in critically ill patients. J Clin Anesth 2002; 14: 210213.
Hanique G, Dugernier T, Laterre PF, et al. Evaluation of oxygen uptake and delivery in critically ill patients: a statistical reappraisal. Intens Care Med 1994; 20: 1926.
Bizouarn P, Blanloeil Y, Pinaud M. Comparison between oxygen consumption calculated by Fick's principle using a continuous thermodilution technique and measured by indirect calorimetry. Br J Anaesth 1995; 75: 719723.
Sakka SG, Reinhart K, Wegscheider K, Meier-Hellmann A. Is the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output? J Cardiothorac Vasc Anesth 2000; 14: 119124.
Caruso LJ, Layon AJ, Gabrielli A. What is the best way to measure cardiac output? Who cares, anyway? Chest 2002; 122: 771774.
Lewis FR, Elings VB, Hill SL, Christensen JM. The measurement of extravascular lung water by thermal-green dye indicator dilution. Ann NY Acad Sci 1982; 384: 394410.
Harris AP, Miller CF, Beattie C, Rosenfeld GI, Rogers MC. The slowing of sinus rhythm during thermodilutioncardiac output determination and the effect of altering injectate temperature. Anesthesiology 1985; 63: 540541.
von Spiegel T, Wietasch G, Bursch J, Hoeft A. Cardiac output determination with transpulmonary thermodilution. An alternative to pulmonary catheterization? Anaesthesist 1996; 45: 10451050.

Keywords

Comparison of cardiac output measurements by arterial trans-cardiopulmonary and pulmonary arterial thermodilution with direct Fick in septic shock

  • G. Marx (a1), T. Schuerholz (a1), R. Sümpelmann (a2), T. Simon (a1) and M. Leuwer (a3)...

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed