Quantitative Analysis of Selected Minor and Trace Elements Through Use of a Computerized Automatic X-Ray Spectrograph

  • B. P. Fabbi (a1), H. N. Elsheimer (a1) and L. F. Espos (a1)

Abstract

Upgrading a manual X-ray spectrograph, interfacing with an 8K computer, and employment of interelement correction programs have resulted in a several-fold increase in productivity for routine quantitative analysis and an accompanying decrease in operator bias both in measurement procedures and in calculations. Factors such as dead time and self-absorption also are now computer corrected, resulting in improved accuracy. All conditions of analysis except for the X-ray tube voltage are controlled by the computer, which enhances precision of analysis. Elemental intensities are corrected for matrix effects, and from these the percent concentrations are calculated and printed via teletype.

Interelement correction programs utilizing multiple linear regression are employed for the determination of the following minor and trace elements; K, S, Rb, Sr, Y, and Zr in silicate rocks, and Ba, As, Sb, and Zn in both silicate and carbonate rock samples. The last named elements use the same regression curves for both rock types. All these elements are determined in concentrations generally ranging from 0.0025% to 4.00%, The sensitivities obtainable range from 0.0001% for barium to 0.001% for antimony. The accuracy, as measured by the percent relative error for a variety of silicate and carbonate rocks, is on the order of 1-7%. The exception is yttrium (13.8%).

Copyright

References

Hide All
1. Sherman, J., “The Theoretical Derivation of Fluorescent X-Ray Intensities from Mixtures,” Spectrochim, Acta, 7, 283306 (1955).
2. Sherman, J., “Simplification of a Formula in the Correlation of Fluorescent X-Ray Intensities from Mixtures,” Spectrochim. Acta, 15, 466470 (1959).
3. Beattie, H. J. and Brissey, R. M., “Calibration Method for X-Ray Fluorescence Spectrometry,” Anal, Chem., 26, 980-983 (1954).
4. Marti, W., “Determination of the Interelement Effect in the X-Ray Fluorescence Analysis of Chromium in Steels,” Spectrochim. Acta, 17, 379383 (1961).
5. Marti, W., “On the Determination of the Interelement Effect in the X-Ray Fluorescence Analysis of Steels,” Spectrochim. Acta, 18, 1499-1504 (1962).
6. Traill, R. J. and Lachance, G. R., “A New Approach to X-Ray Spectrochemical Analysis,” Geol. Surv. Can, Paper, 64-57, 112 (1964).
7. Lachance, G. R. and Traill, R. J., “A Practical Solution to the Matrix Problem in X-Ray Analysis,” Can. Spectrosc., 11, 4348 (1966).
8. Traill, R. J. and Lachance, G. R., “A Practical Solution to the Matrix Problem in X-Ray Analysis. II, Application to a Multicomponent Alloy System,” Can. Spectrosc., 11, 6371 (1966).
9. Claisse, F. and Quintin, M., “Generalization of the Lachance-Traill Method for the Correction of the Matrix Effect in X-Ray Fluorescence Analysis,” Can. Spectrosc., 12, 129-133 (1967).
10. Rousseau, R. and Claisse, F., “Theoretical Alpha Coefficients for the Claisse-Quintin Relation for X-Ray Spectrochemical Analysis,” X-Ray Spectrom., 3, 3136 (1974).
11. Lucas-Tooth, J. and Pyne, C., “The Accurate Determination of Major Constituents by X-Ray Fluorescent Analysis in the Presence of Large Interelement Effects,” in Henke, B. L., Newkirk, J. B. and Mallett, G. R., Bditors, Advances in X-Ray Analysis, Vol. 7, 523-541 Plenum Press (1964).

Quantitative Analysis of Selected Minor and Trace Elements Through Use of a Computerized Automatic X-Ray Spectrograph

  • B. P. Fabbi (a1), H. N. Elsheimer (a1) and L. F. Espos (a1)

Metrics

Full text views

Total number of HTML views: 0
Total number of PDF views: 0 *
Loading metrics...

Abstract views

Total abstract views: 0 *
Loading metrics...

* Views captured on Cambridge Core between <date>. This data will be updated every 24 hours.

Usage data cannot currently be displayed